Z-1321-2022Class II

ABL800 Flex Analyzer Model Numbers 393-800 and 393-801

Radiometer Medical ApS / initiated 2022-05-18 / Open, Classified / root cause: software

Radiometer Medical ApS began a recall of ABL800 Flex Analyzer Model Numbers 393-800 and 393-801 on . FDA classified it as Class II and gives the reason as "There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1321-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Radiometer Medical ApS
Product
ABL800 Flex Analyzer Model Numbers 393-800 and 393-801
Product code
CHL Electrode Measurement, Blood-Gases (Pco2, Po2) And Blood Ph
Reason for recall
There is a potential for sporadic incidents of positive and negative biases for analyzer systems configured with cNa+, cCa+, and cK+.
Root cause tag
software
FDA root cause
Software design
Action
On May 20, 2022, Radiometer issued a "Urgent: Medical Device Recall" notification to affected consignees via: Email and FedEx 2nd day Mail. Radiometer ask consignees to take the following actions: 1. For the affected micro measuring modes Radiometer kindly requests you to not use results for cNa+. The most efficient way to ensure this is to remove the parameter from the affected modes as per the procedures included within the letter. 2. Please complete the Recall Response Form and return to your Radiometer representative ([email protected]).
Quantity in commerce
1489 analyzers
Distribution
Worldwide distribution - US Nationwide distribution in the states of AR, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of Bermuda, Canada, Italy, Norway, Austria, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, China, Croatia, Denmark, Iceland, Germany, Spain, Portugal, India, Brunei, Japan, United Kingdom, Ireland, Russia, Hungary, Australia, New Zealand, Tokelau, Switzerland, France, Czech Republic, Slovakia, South Africa, Botswana, Namibia, Poland, Turkey, Sweden, Finland, Estonia, Kazakhstan, Malaysia, Hong Kong, Mexico, South Korea, Singapore.
Product codes and lots
All ABL800 Flex Analyzers containing software versions below V6.19 MR2; UDI/DI ABL810, 820, and 830, ABL805, ABL815, ABL825, ABL835: 05700693938004 ; UDI/DI ABL817, ABL827, ABL837: 05700693938011

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1321-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1321-2022Class IIOngoingVoluntary: Firm initiated