Z-1297-2022Class II

Atrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the...

Atrium Medical Corporation / initiated 2022-04-21 / Open, Classified / root cause: labeling and UDI

Atrium Medical Corporation began a recall of Atrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the chest cavity or mediastinum. Product REF Number... on . FDA classified it as Class II and gives the reason as "Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 () while the case and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1297-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Atrium Medical Corporation
Product
Atrium Firm PVC Thoracic Catheter 28French-Intended to facilitate the evacuation of air and/or fluid from the chest cavity or mediastinum. Product REF Number: 15028
Product code
GBZ Catheter, Cholangiography
Reason for recall
Mismatch between the expiration date on the sterile barrier pouch and the expiration date on the case/shipper label. Product sterile barrier pouch label expiration date of 16SEP2024 (2024-09-16) while the case and shipper labels both showed 21SEP2024 (2024-09-21), causing product being used for up to five (5) days past the true expiration date
Root cause tag
labeling and UDI
FDA root cause
Under Investigation by firm
Action
Getinge issued Urgent Medical Device Removal to consignees via FedEx with Signature Proof of Delivery on 4/21/22. Letter states reason for recall, health risk and action to take: Please examine your inventory immediately to determine if you have any of the Atrium Firm PVC Thoracic Catheter 28French with the REF/LOT number listed in this notice. Should you have any affected product, please remove from area(s) of use. The LOT Number (ME231010) can be found on the product label illustrated. " If you have affected product, you are entitled to a replacement at no cost to your facility. You will receive a replacement upon your acknowledgement that you have affected product for return. " Please forward this information to all current and potential thoracic cathether users within your hospital / facility. If you are a distributor who has shipped any affected products to customers, please forward this document to their attention for appropriate action. Please contact your local Atrium/Getinge Customer Service department to request a return authorization (RMA) and shipping instructions to return any affected product. Whether or not you have affected product(s) with the REF/LOT numbers listed in this notice, please complete and sign the attached MEDICAL DEVICE- REMOVAL RESPONSE FORM (page 4) to acknowledge that you have received this notification. Return the completed form to Getinge by e-mailing a scanned copy to [email protected] or by faxing the form to (877) 435-7043. Contact your Atrium/Getinge representative or call the Atrium/Getinge Customer Support at (603) 880-1433 (press option 1), Monday through Friday, between the hours of 8:00 a.m. and 6:00 p.m. (Eastern Time Zone).
Quantity in commerce
56 cases (560 units / 10 units per case)
Distribution
AZ IL KS LA MI NY OH OK TX VA
Product codes and lots
UDI-DI: 20650862150289 - case / 00650862150285- each Lot Number: ME231010

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1297-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1297-2022Class IIOngoingVoluntary: Firm initiated