Z-1276-2022Class I
Clean Catch Urine Kit NS 178489, AHMU58A
American Contract Systems, Inc. / initiated 2022-05-02 / Open, Classified / root cause: manufacturing process
American Contract Systems, Inc. began a recall of Clean Catch Urine Kit NS 178489, AHMU58A on . FDA classified it as Class I and gives the reason as "Affected products were assembled off-site by individuals who may not have been properly trained". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1276-2022
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- Clean Catch Urine Kit NS 178489, AHMU58A
- Product code
- JSM Culture Media, Non-Propagating Transport
- Reason for recall
- Affected products were assembled off-site by individuals who may not have been properly trained.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On May 2, 2022, the customer was notified by email. The customer was instructed to destroy or discard affected kits.
- Quantity in commerce
- 274,400 kits total
- Distribution
- US distribution in the state of Minnesota.
- Product codes and lots
- UDI-DI: 00191072120557 Lots 891211, 737211, 699211, 986221
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1276-2022 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-1277-2022 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY