SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber...
Olympus Corporation of the Americas began a recall of SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX200S (Pouch) - 200 Micron... on . FDA classified it as Class II and gives the reason as "Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1296-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- OLYMPUS Corporation of the Americas
- Product
- SOLTIVE SuperPulsed Laser Fibers, Model# Labeled TFL-FBX150BS (Carton) - 150 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX200S (Pouch) - 200 Micron Single Use Fiber Model # Labeled TFL-FBX200BS (Carton) - 200 Micron Ball Tip Single Use Fiber Model # Labeled TFL-FBX550S (Carton) - 550 Micron Single Use Fiber Model # Labeled TFL-FBX200S (Carton) - 200 Micron Single Use Fiber Model # Labeled TFL-FBX365S (Pouch) - 200 Micron Single Use Fiber
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Device model and size for affected lots of Laser Fibers may be mismatched between the product's carton, immediate pouch packaging, and/or the device.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling mix-ups
- Action
- Firm notified customers by letter on May 26, 2022. Customers were instructed to return on hand inventory to Olympus for credit.
- Quantity in commerce
- 190 cartons (950 pieces total)
- Distribution
- Nationwide distribution in the US. Foreign distribution worldwide.
- Product codes and lots
- UDI-DI Lot Model# Labeled 00821925043923 KR149856 TFL-FBX150BS (Carton) 00821925043923 KR149856 TFL-FBX200S (Pouch) 00821925043985 KR195775 TFL-FBX200BS (Carton) 00821925044043 KR197789 TFL-FBX550S (Carton) 00821925043930 KR206254 TFL-FBX200S (Carton) 00821925043930 KR206254 TFL-FBX365S (Pouch)
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1296-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY