NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System...

Synapse Biomedical Inc / initiated 2022-05-31 / Terminated / root cause: labeling and UDI

Synapse Biomedical Inc began a recall of NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System Kit 20-0036 NeuRx Diaphragm Pacing System, Patient... on . FDA classified it as Class III and gives the reason as "Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport...". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1393-2022
Classification
Class III
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Synapse Biomedical Inc
Product
NeuRx Diaphragm Pacing System (NeuRx DPS), Affected part numbers: Part # Part Name 20-0035 NeuRx DPS System Kit 20-0036 NeuRx Diaphragm Pacing System, Patient Kit, SCI 20-0028 NeuRx DPS Surgical Kit, HDE 20-0040 NeuRx Diaphragm Pacing System Spare Sterile Components 21-0001-99 Mapping Probe 21-0004-99 Indifferent Electrode 21-0007-99 Lead Tunneler Set 21-0009-99 Blue Cable Set 21-0010-99 Transducer Tubing 21-0017-99 PermaLoc Electrode 21-0023-99 Green Cable 21-0024-99 Orange Cable 21-0025-99 Yellow Cable 21-0026-99 White Cable 77-0050 Surgeon Instruction Manual 20-0045 NeuRx DPS System Kit, ALS HDE 20-0046 NeuRx Diaphragm Pacing System, Patient Kit, ALS Marketed under HDE H070003 as well as H100006
Product code
OIR Diaphragmatic/Phrenic Nerve Laparoscopically-Implanted Stimulator
Reason for recall
Firm has revised labeling to consistently call out storage conditions and revised the instructions for use to clarify that the temperature range symbol is intended to indicate storage conditions (versus transport conditions or tolerance conditions for the package material).
Root cause tag
labeling and UDI
FDA root cause
Labeling design
Action
On May 31, 2022, the firm began to contact all affected customer sites. Customers were informed that the way the temperature storage conditions are stated on the labeling of these devices could lead to improper storage that, under certain circumstances, could damage the devices. Synapse will visit customer facilities to place a label with proper storage conditions over the current storage conditions on the device label. Customers who would rather not have a site visit may destroy their devices and Synapse will replace them. The firm will also provide a revised Surgeon Instruction Manual that includes a description of the storage temperature symbol.
Quantity in commerce
71
Distribution
Domestic distribution to AL CA CO FL GA IL MA MI MN NC NJ NY OH OK OR PA SC TX WV. Foreign distribution pending.
Product codes and lots
UDI-DI: 00852184003007 00852184003045 00852184003038 00852184003267 00852184003090 00852184003106 00852184003113 00852184003120 00852184003137 00852184003144 00852184003151 00852184003168 00852184003175 00852184003182 00852184003281 00852184003052 Lots/Serial Numbers: 20-0035-113021-3-3 20-0035-120221-2-1 20-0035-121621-5-2 20-0035-121621-5-3 20-0035-010722-5-4 20-0035-010722-5-5 20-0035-030722-2-2 20-0035-031122-5-1 20-0035-031122-5-2 20-0035-031122-5-5 20-0035-033122-5-1 20-0035-033122-5-2 20-0035-033122-5-3 20-0035-040622-4-1 20-0035-040622-4-2 20-0035-040622-4-3 20-0035-040622-4-2 20-0035-042122-1-1 20-0028-060221-5 20-0028-062421-5 20-0028-062521-5 20-0028-072821-5 20-0028-031022-5 20-0040-012221-10 20-0040-012221-10 20-0040-021921-9 20-0040-012221-10 20-0040-012221-10 20-0040-021921-9 20-0040-021921-9 20-0040-021921-9 20-0040-021921-9 20-0040-012621-10 20-0040-012621-10 20-0040-012621-10 20-0040-012621-10 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060221-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-060821-5 20-0040-071321-6 20-0040-071321-6 20-0040-092021-10 20-0040-092021-10 20-0040-071521-2 20-0040-092021-10 20-0040-092021-10 20-0040-092021-10 20-0040-100721-10 20-0040-100721-10 20-0040-100721-10 20-0040-100721-10 20-040-100721-10 20-0040-100721-10 20-0040-100721-10 20-0040-122121-10 20-0040-122121-10 20-0040-122121-10 20-0040-122121-10 20-0040-010622-10 20-0040-121621-1 20-0040-020822-10 20-0040-010622-10 20-0040-040122-2

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1393-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1393-2022Class IIITerminatedVoluntary: Firm initiated