Haiou Needle Retractable Safety Syringe with Needle-1ML 23G x 1 (0.6x25mm). 200pcs/Box
Guangdong Haiou Medical Apparatus Co., Ltd began a recall of Haiou Needle Retractable Safety Syringe with Needle-1ML 23G x 1 (0.6x25mm). 200pcs/Box Intended Use: Disposable sterile needle retractable safety syringe... on . FDA classified it as Class II and gives the reason as "Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1507-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Guangdong Haiou Medical Apparatus Co., Ltd.
- Product
- Haiou Needle Retractable Safety Syringe with Needle-1ML 23G x 1 (0.6x25mm). 200pcs/Box Intended Use: Disposable sterile needle retractable safety syringe. Catalog Number: NRSS1-23G25
- Product code
- MEG Syringe, Antistick
- Reason for recall
- Complaints received in which needle and needle holder have detached from the syringe after injection as well as needle not retracting into the syringe as per intended use, resulting in an increased risk of needle stick injury to the end user
- FDA root cause
- No Marketing Application
- Action
- Guangdong Haiou Medical., LTD (Haiou) contacted direct consignee by email on [10-Jun-2022] and Formal Recall Urgent Medical Device Recall letter through email and first-class letter. Letter states reason for recall, health risk and action to take: " Remove product from sale and cease distribution of product subject to this recall action. " Immediately review your inventory for the specific Product Description and Lot Number listed above and place into quarantine until further notice. " Notify us of any stock you have still under your direct control " Notify all customers to whom you have distributed this product of this recall. Use the template provided with this letter. " Complete the attached Customer Response Form and return to the Haiou contact noted on the form whether or not you have any of the impacted material so that Haiou may acknowledge your receipt of this notification per FDA requirements and subsequently process your product replacement. " Forward any customer complaints for the use of this product to Haiou as per the contact information provided. Jiehong Luo QA Manager Phone:+ 86-663-2900999 Email: [email protected] Address: Nanyuan Industrial Zone, North Liusha, Puning City, Guangdong, China
- Quantity in commerce
- 7,218,000 units
- Distribution
- US Nationwide distribution in the state of CA.
- Product codes and lots
- Lot Number: 20JC2 Exp. Date: 2025/10/17; 21AA0 Exp Date: 2026/01/04
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1506-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1507-2022 | Class II | Ongoing | Voluntary: Firm initiated |