Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray...
Siemens Medical Solutions USA, Inc began a recall of Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System on . FDA classified it as Class II and gives the reason as "It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are...". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1343-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Siemens Medical Solutions USA, Inc.
- Product
- Artis zeego (Model no. 10280959) and Artis Q.zeego (Model no. 10848283), Interventional Fluoroscopic X-Ray System
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- It may occur that after system startup no stand movement is possible any longer in the event of a discharged BIOS battery of the robotic stand control PC. If this problem occurs during startup, all stand movements are blocked and can only be reactivated by a field service engineer.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- On May 24, 2022, the firm mailed Urgent Medical Device Correction letters to affected consignees. Customers were informed that if the system is not used regularly, please ensure that the system is switched on in due time before any use to check the status. Please ensure that patient treatment can be continued in other ways if there is any possible danger for the safety of the patient. The firm will replace the Bios battery and the service interval will be shortened to two years. The firm will contact customers to arrange a date to perform the correction. Customers may call the service organization at 1-800-888-7435 to schedule an earlier appointment.
- Quantity in commerce
- 233 worldwide; 32 US
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CO, FL, IL, LA, ME, MN, MO, NC, OH, PA, SC, TN, TX, WI and the countries of Australia, Austria, Azerbaijan, Belgium, Canada, China, Finland, France, Germany, Hong Kong, India, Italy, Japan, Lebanon, Lithuania, Luxembourg, Netherlands, Poland, Portugal, South Korea, Switzerland, United Kingdom.
- Product codes and lots
- UDI-DI: Artis zeego 04056869010083 Artis Q.zeego 04056869010007
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1343-2022 | Class II | Terminated | Voluntary: Firm initiated |