Z-1367-2022Class II

Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic...

Ortho-Clinical Diagnostics, Inc. / initiated 2022-05-26 / Open, Classified / root cause: labeling and UDI

Ortho-Clinical Diagnostics, Inc. began a recall of Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems... on . FDA classified it as Class II and gives the reason as "On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.

Recall record

Recall number
Z-1367-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems. Product Code: 6199960
Product code
QKO Reagent, Coronavirus Serological
Reason for recall
On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).
Root cause tag
labeling and UDI
FDA root cause
Incorrect or no expiration date
Action
Ortho Clinical Diagnostics issued Important Correction Notification letter (CL2022-135) on May 26, 2022. Letter states reason for recall, health risk and action to take: When using VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack, do not store on-analyzer for longer than 6 weeks. " Until a lot update occurs, OAS for VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack will temporarily require manual monitoring and removal of reagent packs from analyzer no longer than 6 weeks after packs have been opened and loaded onto the analyzer. " Complete the enclosed Confirmation of Receipt form no later than June 9, 2022. " Post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System in your laboratory until further notice. " Please forward this notification if VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311
Quantity in commerce
20858
Distribution
Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.
Product codes and lots
UDI-DI :10758750034413. All expired, in-date and future released lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1367-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1367-2022Class IIOngoingVoluntary: Firm initiated