Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic...
Ortho-Clinical Diagnostics, Inc. began a recall of Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems... on . FDA classified it as Class II and gives the reason as "On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1367-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ortho-Clinical Diagnostics, Inc.
- Product
- Anti-SARS-CoV-2 IgG Quantitative Reagent Pack- Intended for use on VITROS ECi/ECiQ/3600 Immunodiagnostic Systems and VITROS 5600/XT 7600 Integrated Systems. Product Code: 6199960
- Product code
- QKO Reagent, Coronavirus Serological
- Reason for recall
- On-Analyzer Stability (OAS) time, supported for 6 weeks; not 8 weeks as is currently stated in the Instructions For Use (IFU).
- Root cause tag
- labeling and UDI
- FDA root cause
- Incorrect or no expiration date
- Action
- Ortho Clinical Diagnostics issued Important Correction Notification letter (CL2022-135) on May 26, 2022. Letter states reason for recall, health risk and action to take: When using VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack, do not store on-analyzer for longer than 6 weeks. " Until a lot update occurs, OAS for VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack will temporarily require manual monitoring and removal of reagent packs from analyzer no longer than 6 weeks after packs have been opened and loaded onto the analyzer. " Complete the enclosed Confirmation of Receipt form no later than June 9, 2022. " Post this notification by each VITROS ECi/ECiQ/3600/5600/XT 7600 System in your laboratory until further notice. " Please forward this notification if VITROS Anti-SARS-CoV-2 IgG Quantitative Reagent Pack was distributed outside of your facility. If you have further questions, please contact Ortho Care Technical Solutions Center at 1-800-421-3311
- Quantity in commerce
- 20858
- Distribution
- Worldwide distribution - US Nationwide and the countries of Colombia, India, Singapore.
- Product codes and lots
- UDI-DI :10758750034413. All expired, in-date and future released lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1367-2022 | Class II | Ongoing | Voluntary: Firm initiated |