Z-1344-2022Class II

Abbott TactiCath Sensor Enabled, Contact Force Ablation Catheter, 8F 115cm FJ, REF A-TCSE-FJ. For mapping of...

Abbott Medical / initiated 2022-06-09 / Open, Classified / root cause: design

Abbott began a recall of Abbott TactiCath Sensor Enabled, Contact Force Ablation Catheter, 8F 115cm FJ, REF A-TCSE-FJ. For mapping of the heart chambers during ablation on . FDA classified it as Class II and gives the reason as "When connected to the EnSite Precision Navigation System, an affected TactiCath Contact Force Ablation Catheter, Sensor Enabled may present the error message: "invalid catheter,, or "expired catheter."". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1344-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Abbott TactiCath Sensor Enabled, Contact Force Ablation Catheter, 8F 115cm FJ, REF A-TCSE-FJ. For mapping of the heart chambers during ablation.
Product code
OAE Catheter, Percutaneous, Cardiac Ablation, For Treatment Of Atrial Fibrillation
Reason for recall
When connected to the EnSite Precision Navigation System, an affected TactiCath Contact Force Ablation Catheter, Sensor Enabled may present the error message: "invalid catheter,, or "expired catheter."
Root cause tag
design
FDA root cause
Device Design
Action
The firm notified its consignees by hand delivered Medical Device Recall letter on 06/09/2022. The letter explained the problem and provided the following instructions: "Next Steps To help reduce risk, we recommend: " Do not use any remaining inventory from the affected lots listed in Appendix A. " Complete and return the accompanying Acknowledgement Form to Abbott. " Return all remaining unused affected devices to Abbott. Your Abbott representative can assist you in returning these devices and obtaining replacements. Please forward this notice to anyone within your organization who may need to be notified and maintain a record of this notice along with a copy of the completed Acknowledgement Form for your records."
Quantity in commerce
191 devices
Distribution
US Nationwide distribution.
Product codes and lots
UDI 01)05415067027665(17)230831(10)xxxxxxx, Batch Numbers 8409416, 8409675, 8410811

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1344-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1344-2022Class IITerminatedVoluntary: Firm initiated