NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3
Fresenius Medical Care Holdings, Inc. began a recall of NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3 on . FDA classified it as Class II and gives the reason as "The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream...". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.
Recall record
- Recall number
- Z-1311-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- NaturaLyte 2301 2K 3.0 Ca 1 Mg Gal, Model No. 08-2301-3
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- The product was potentially exposed to below-recommended storage temperatures, which may cause leaks in the packaging. Leaks can lead to biological contamination and therefore an increased risk of potential bloodstream infection.
- Root cause tag
- packaging
- FDA root cause
- Under Investigation by firm
- Action
- On June 1, 2022, the firm reached out to its affected customers via telephone to explain the product issue. Customers were instructed to identify and segregate the affected product immediately and remove the product from inventory. If customers had any product to return, the recalling firm would have a customer service representative follow up with the customer. Customers were reminded to always inspect packaging for leaks.
- Quantity in commerce
- 3 cases
- Distribution
- US Nationwide distribution in the states of Texas and Louisiana.
- Product codes and lots
- UDI: (01)00840861101702 (17) 230131 (10) 21AGAC009 Lot 21AGAC009
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1311-2022 | Class II | Terminated | Voluntary: Firm initiated |