Z-1427-2022Class II

GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia...

GE Healthcare / initiated 2022-05-27 / Open, Classified / root cause: manufacturing process

GE Healthcare, LLC began a recall of GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000 on . FDA classified it as Class II and gives the reason as "The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1427-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Healthcare
Product
GE Healthcare Avance CS2, Avance CS2 Pro Anesthesia Systems, Part #1009-9050-000; and Aisys CS2 Anesthesia Systems, Part # 1011-9050-000.
Product code
BSZ Gas-Machine, Anesthesia
Reason for recall
The base of the anesthesia system can have a crack and could potentially fracture resulting in a tip or overbalance of the anesthesia system. This can result in potential injury if it falls on a person.
Root cause tag
manufacturing process
FDA root cause
Process control
Action
The recalling firm issued letters dated 5/27/2022 via traceable means on 5/27/2022 which were addressed to the Director of Respiratory, Chief of Anesthesia, Health Care Administrator/Risk Manager, and Director of Biomedical/Clinical Engineering. The letter is flagged that the consignee needs to ensure all potential users in the facility are made aware of this safety notification and recommended actions. The notification informs the consignees of the affected devices, the safety issue, and actions to be taken. The actions notified the consignee they can continue to use the anesthesia system by following instructions listed in the letter. The instructions were to inspect the anesthesia system for cracks at the rear of the base as depicted in the diagram and photograph, and if no cracks are observed, they can continue to use the anesthesia system. If a crack at the rear is observed, they can continue to use the system under the following conditions: Limit the movement of the system using care on uneven flooring/threshold and ensure the floor is clear of obstacles (cables, power cords, etc.); and do not exceed the recommended weight limit for equipment mounted or supported by the anesthesia system. A Medical Device Notification Acknowledgement Response form was to be completed and returned within 30 days of receipt to acknowledge receipt and understanding of the notification. The letter informs the consignee that the recalling firm will inspect all affected devices at no cost and that a firm representative will contact the consignee to arrange for the correction.
Quantity in commerce
667 devices
Distribution
Distribution was made to AZ, CA, CO, CT, FL, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MO, MT, NC, NE, NV, NY, OH, PA, SC, TN, TX, VA, and WV. There was also military distribution but no government distribution. Foreign distribution was made to Australia, Canada, China, Estonia, Finland, France, Germany, Greece, India, Indonesia, Iraq, Ireland, Italy, Japan, Jordan, Kenya, Korea, Malaysia, New Zealand, Philippines, Romania, Russia, Saudi Arabia, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Vietnam.
Product codes and lots
Avance CS2 and Avance CS2 Pro - GTIN 00840682102322, all serial numbers; Aisys CS2 - GTIN 00840682102292, all serial numbers.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1427-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1427-2022Class IIOngoingVoluntary: Firm initiated