Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D)
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) began a recall of Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D) on . FDA classified it as Class II and gives the reason as "Manufacturing error that may have resulted in a cathode component being out of specification. All devices met final functional testing requirements". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1480-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Cobalt Implantable Cardioverter Defibrillators with Cardiac Resynchronization Therapy (CRT-D)
- Product code
- NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)
- Reason for recall
- Manufacturing error that may have resulted in a cathode component being out of specification. All devices met final functional testing requirements.
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- On 05-Nov-2020, Medtronic initiated retrieval activities via verbal communication to consignees who were in possession of unused, affected devices according to Medtronic records. Consignees were advised to quarantine all affected devices within their possession. In the United States, a written communication was mailed to two (2) identified consignees on 24-Nov-2020, directing the return of all unused, affected devices.
- Quantity in commerce
- 3 units
- Distribution
- US Nationwide distribution in the states of Indiana and CA.
- Product codes and lots
- GTIN 00763000178178, Serial Numbers: RTK601865S, RTK601825S, RTK601828S
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1480-2022 | Class II | Ongoing | Voluntary: Firm initiated |