Z-1374-2022Class II
Ventana HE 600 System, automated slide preparer, for use in laboratories
Ventana Medical Systems Inc / initiated 2022-06-01 / Open, Classified
Ventana Medical Systems Inc began a recall of Ventana HE 600 System, automated slide preparer, for use in laboratories on . FDA classified it as Class II and gives the reason as "There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1374-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ventana Medical Systems Inc
- Product
- Ventana HE 600 System, automated slide preparer, for use in laboratories.
- Product code
- KPA Slide Stainer, Automated
- Reason for recall
- There is a potential for fluid leak inside the Ventana HE 600 instrument which could result in electrical short and fire
- FDA root cause
- Under Investigation by firm
- Action
- On June 6, 2022 Ventana Medical Systems, Inc. (via Roche Diagnostics Corp.) issues an "Urgent Medical Device Correction" notification to affected consignees. Ventana Medical ask consignees to take the following actions: 1. Notify all operators of this potential hazard and forward this UMDC as appropriate. If your facility has distributed the affected device to another site, please ensure this UMDC is provided to that site. 2. Do not leave the instrument unattended when the power is on. Turn off the power to the instrument when it is unattended per the instructions in this UMDC. 3. Do not place the system in stand-by mode overnight when unattended until further mitigation actions have been implemented. 4. If fire is detected, follow your local emergency guidelines. 5. If a leak, smoke, or odor is detected in your VENTANA HE 600 instrument, power off the instrument immediately and call the Roche Support Network Customer Support Center, 24 hours a day, seven days a week at 1-800-227-2155. 6. Complete the enclosed faxback form (TP-01573) and return it according to the instructions on the form. 7. File this Urgent Medical Device Correction (UMDC) for future reference. 8. Adverse events or quality problems experienced with the use of this product may also be reported to the FDA s MedWatch Adverse Events Reporting Program: Online at http://www.fda.gov/Safety/MedWatch/HowToReport/default.htm (form available to fax or mail), or call FDA 1-800-FDA-1088.
- Quantity in commerce
- 763 instruments
- Distribution
- US: AL AR AZ CA CO CT DC FL GA HI IA IL IN KS KY MA MD MI MN MO MS MT NC NE NJ NY OH OK OR PA PR SC SD TN TX UT VA WA WI WV OUS: Canada, Austria, Belgium, Croatia, Denmark, Finland, France, Germany, Greece, Ireland, Italy, Luxembourg, Netherlands, Norway, Poland, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, UK, Argentina, Australia, Brazil, Chile, China, Colombia, Costa Rica, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Israel, Japan, Kuwait, Lebanon, Malaysia, Saudi Arabia, Singapore, South Korea, Taiwan, Thailand, United Arab Emirates (UAE), Vietnam
- Product codes and lots
- All Lots; UDI/DI: 04015630976010
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1374-2022 | Class II | Ongoing | Voluntary: Firm initiated |