Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1468-2022Class II

HammerTube Implant, 2.75mm, 10 Cannulated (Sterile)

Paragon 28 / initiated 2022-05-26 / Open, Classified

Paragon 28, Inc. began a recall of HammerTube Implant, 2.75mm, 10 Cannulated (Sterile) on . FDA classified it as Class II and gives the reason as "Due to Titanium plasma coating not present implant". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1468-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Paragon 28
Product
HammerTube Implant, 2.75mm, 10 Cannulated (Sterile)
Product code
HTY Pin, Fixation, Smooth
Reason for recall
Due to Titanium plasma coating not present implant.
FDA root cause
Under Investigation by firm
Action
On 05/26/2022, the firm communicated via email to its customers/distributors informing them that Paragon 28 has identified via customer complaints that the titanium plasma coating is missing on the HammerTube Implants. On 06/15/2022, Paragon28 communicated an updated "URGENT Medical Device Recall" letter to customer that informed them of the potential health risks involved with the affected devices such as: " The defect may cause the implant to potentially result in reduced of pull-out force creating instability at the joint which may lead to delayed or lack of healing. " The implants, with the defect, may cause potential mating interaction failure between inserter and implant. " The defect may lead to an incorrect intended implant placement; directional and or depth. Customers are instructed to: 1. Review physical inventory, discontinue use, quarantine and not to further distribute the affected devices. 2. Return affected devices immediately to: Paragon28, Attn: Recalls 14445 Grasslands Dr. Englewood, CO 80112 Any questions, contact: Mandy Kinnee Email: [email protected] Phone: 720.966.8333 *Note: You may also contact [email protected]. 1.
Quantity in commerce
9 implants
Distribution
U.S. Nationwide distribution in the states of AL, CA, GA, IL and KY.
Product codes and lots
Part Number: P40-002-0275-S UDI-DI Code: (01)00889795033987(17)270323(10)26010822202 Lot Number: 26010822202

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1468-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1468-2022Class IIOngoingVoluntary: Firm initiated