Z-1468-2022Class II
HammerTube Implant, 2.75mm, 10 Cannulated (Sterile)
Paragon 28 / initiated 2022-05-26 / Open, Classified
Paragon 28, Inc. began a recall of HammerTube Implant, 2.75mm, 10 Cannulated (Sterile) on . FDA classified it as Class II and gives the reason as "Due to Titanium plasma coating not present implant". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1468-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Paragon 28
- Product
- HammerTube Implant, 2.75mm, 10 Cannulated (Sterile)
- Product code
- HTY Pin, Fixation, Smooth
- Reason for recall
- Due to Titanium plasma coating not present implant.
- FDA root cause
- Under Investigation by firm
- Action
- On 05/26/2022, the firm communicated via email to its customers/distributors informing them that Paragon 28 has identified via customer complaints that the titanium plasma coating is missing on the HammerTube Implants. On 06/15/2022, Paragon28 communicated an updated "URGENT Medical Device Recall" letter to customer that informed them of the potential health risks involved with the affected devices such as: " The defect may cause the implant to potentially result in reduced of pull-out force creating instability at the joint which may lead to delayed or lack of healing. " The implants, with the defect, may cause potential mating interaction failure between inserter and implant. " The defect may lead to an incorrect intended implant placement; directional and or depth. Customers are instructed to: 1. Review physical inventory, discontinue use, quarantine and not to further distribute the affected devices. 2. Return affected devices immediately to: Paragon28, Attn: Recalls 14445 Grasslands Dr. Englewood, CO 80112 Any questions, contact: Mandy Kinnee Email: [email protected] Phone: 720.966.8333 *Note: You may also contact [email protected]. 1.
- Quantity in commerce
- 9 implants
- Distribution
- U.S. Nationwide distribution in the states of AL, CA, GA, IL and KY.
- Product codes and lots
- Part Number: P40-002-0275-S UDI-DI Code: (01)00889795033987(17)270323(10)26010822202 Lot Number: 26010822202
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1468-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.