NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate...
Fresenius Medical Care Holdings, Inc. began a recall of NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure Part... on . FDA classified it as Class II and gives the reason as "Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1371-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Fresenius Medical Care Renal Therapies Group, LLC
- Product
- NATURALYTE Liquid Acid 3.0K 3.0Ca- indicated for use in patients undergoing extracorporeal bicarbonate hemodialysis for acute and chronic renal failure Part Number: 08-3301-2
- Product code
- KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
- Reason for recall
- Alternate Bottle Caps labeled with "Sealed for your Protection" on the foil not qualified for use
- Root cause tag
- labeling and UDI
- FDA root cause
- Under Investigation by firm
- Action
- Fresenius Medical Care issued Urgent Medical Device Recall letter (FA-2022-11-T) dated 6/2/22. Letter states reason for recall, health risk and action to take: These instructions apply only to clinics directly contacted by the Fresenius Medical Care Corrections and Removals Department. " Immediately examine your stock to determine whether you have NaturaLyte Liquid Acid concentrate with the part and lot numbers listed below. " Only bottles with sealed for your protection on the bottle cap foil are affected (see picture below). " Remove the affected product from your inventory. " FMCRTG, LLC will contact you regarding product return. Please contact FMCRTG Customer Service at 1-800-323-5188 if you have any further questions.
- Quantity in commerce
- 163 cases
- Distribution
- Natiowide
- Product codes and lots
- UDI-DI: (01) 10840861101757 (17) 240331 (10) 22CTAC102 Lot Number: 22CTAC102 Exp Date: 2024-03-31
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1370-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1371-2022 | Class II | Ongoing | Voluntary: Firm initiated |