GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and...
GE Healthcare, LLC began a recall of GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data within a computer network or on a workstation on . FDA classified it as Class II and gives the reason as "Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1333-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Healthcare
- Product
- GE Centricity Universal Viewer Zero Footprint. To view, communicate, process, and display Medical images and data within a computer network or on a workstation.
- Product code
- LLZ System, Image Processing, Radiological
- Reason for recall
- Potential to display inaccurate measurements on images in Centricity Universal Viewer Zero Footprint Client (ZFP)
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- GE notified their consignees by issuing an URGENT MEDICAL DEVICE CORRECTION notice on 08 June 2022. The letter explains the two safety issues and provides the following actions be taken: Issue #1: "You can continue to use your device in accordance with the User Manual and the actions below: Do not utilize any measurements when the 'Lossy' overlay is displayed in the lower left of the viewport and, when you hover over the word 'Lossy', it displays the tooltip box with the phrase 'Image scaled down from its original resolution'." Issue #2: "You can continue to use your device in accordance with the User Manuals and the actions below: Do not perform any measurement on non-squared pixel images. Always confirm critical diagnostic measurements by other means (i.e using a fiducial marker of known size)." GE Healthcare will correct all affected products at no cost to you. A GE Healthcare representative will contact you to arrange for the correction. After the software has been corrected, be sure to destroy the installation media for affected software at your site. If you have any questions or concerns regarding this notification, please contact GE Healthcare Service at 1-800-437-1171 or your local Service Representative.
- Quantity in commerce
- 1,934 units
- Distribution
- Worldwide distribution.
- Product codes and lots
- UDI 00840682102988, Versions 6.0 SP6 through SP11.4
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1333-2022 | Class II | Ongoing | Voluntary: Firm initiated |