GreenLight HPSEA Laser Fibers, UPN 0010-2092
Boston Scientific Corporation began a recall of GreenLight HPSEA Laser Fibers, UPN 0010-2092 on . FDA classified it as Class III and gives the reason as "Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1392-2022
- Classification
- Class III
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Boston Scientific Corporation
- Product
- GreenLight HPSEA Laser Fibers, UPN 0010-2092
- Product code
- GEX Powered Laser Surgical Instrument
- Reason for recall
- Several complaints were received for the affected lot; users received a "Fiber type does not match card type" error message on the GreenLight XPS Laser console, necessitating the replacement of the laser fiber and fiber card. This may result in prolonged procedure while the fiber is being replaced.
- Root cause tag
- software
- FDA root cause
- Software Manufacturing/Software Deployment
- Action
- On May 31, 2022, the firm distributed Urgent Medical Device Product Removal letters to affected customers. Customers were instructed to immediately segregate and return the product to Boston Scientific.
- Quantity in commerce
- 57
- Distribution
- Domestic distribution to AR CA IL ME MI NC TX Foreign distribution to Australia, Canada, and France.
- Product codes and lots
- UDI-DI (GTIN): 00878953005522 Lot: 27770466
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1392-2022 | Class III | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY