ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee System. instrument...
Aesculap Implant Systems LLC began a recall of ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee System. instrument used during orthopedic implant surgery on . FDA classified it as Class II and gives the reason as "Nonfunctional key due to the potential for damage to the key pins cause by force of use during assembly of the tibia locking ring". As of , the recall status on record is Completed, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1391-2022
- Classification
- Class II
- Status
- Completed
- Initiated
- Posted
- Recalling firm
- Aesculap Inc
- Product
- ENDURO SPECIAL KEY TIB.LOCK.RING F1/10MM product code NP462R used with EnduRo Total Knee System. instrument used during orthopedic implant surgery.
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Nonfunctional key due to the potential for damage to the key pins cause by force of use during assembly of the tibia locking ring.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- A notification letter was issued November 12, 2019 mailed USPS certified with registered return receipt. The letter identified item number NP462R EnduRo Special Key for Tibia Locking Ring F1 10/12mm as being removed from the market for a new version NP462RM after overall system modification. Users are instructed to 1) Review inventory and check that only the latest version (NP462RM and NP454R) are being used. 2) Contact customer service to return any affected inventory. 3) Complete and return the acknowledgement form. Questions are to be directed to Josie McBride at (610) 984-9290.
- Quantity in commerce
- 67
- Distribution
- US Nationwide distribution in the states of AL, AZ, CA, CO, ID, IL, LA, MA, MO, NC, OK, TX, WI.
- Product codes and lots
- UDI-DI (GTIN): 04046963419639
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1391-2022 | Class II | Completed | Voluntary: Firm initiated |