Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1378-2022Class II

Endo Model Modular Femur and Tibia, PorEx. orthopedic prosthesis. Model Nos.: X-Small, Right 15-3815/11...

Waldemar Link GmbH & Co. KG / initiated 2022-05-24 / Open, Classified / root cause: manufacturing process

Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Endo Model Modular Femur and Tibia, PorEx. orthopedic prosthesis. Model Nos.: X-Small, Right 15-3815/11, Small, Right 15-3816/11, Small, Left 15-3816/12... on . FDA classified it as Class II and gives the reason as "There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1378-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Waldemar Link GmbH & Co. KG
Product
Endo Model Modular Femur and Tibia, PorEx. orthopedic prosthesis. Model Nos.: X-Small, Right 15-3815/11, Small, Right 15-3816/11, Small, Left 15-3816/12, Medium, Right 15-3817/11, Medium, Left 15-3817/12, Large, Right 15-3818/11.
Product code
JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
Reason for recall
There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
On May 24, 2022, the firm notified affected customers via Urgent Field Safety Notice email. Customers were provided a modified surgical technique providing information for the cementing technique for Tilastan proximal spacers to minimize the risk. No product return is required.
Quantity in commerce
3785
Distribution
Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
Product codes and lots
UDI-DI  04026575034727 X-Small, Right; 04026575034741 Small, Right; 04026575034758 Small, Left; 04026575164042 Medium, Right; 04026575164059 Medium, Left; 04026575164028 Large, Right;   Serial/Lot Numbers: 200818/1709 200818/1720 201013/0291 200513/2160 200818/1732 200623/4154 200812/1511 210302/2440

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1378-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1377-2022Class IIOngoingVoluntary: Firm initiated
Z-1378-2022Class IIOngoingVoluntary: Firm initiated
Z-1379-2022Class IIOngoingVoluntary: Firm initiated
Z-1380-2022Class IIOngoingVoluntary: Firm initiated