Tibial Component. orthopedic prosthesis. Model Nos: Small, W 16-2817/02 Medium, W 16-2817/05 Large, W...
Waldemar Link GmbH & Co. KG (Mfg Site) began a recall of Tibial Component. orthopedic prosthesis. Model Nos: Small, W 16-2817/02 Medium, W 16-2817/05 Large, W 16-2817/07 on . FDA classified it as Class II and gives the reason as "There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1379-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Waldemar Link GmbH & Co. KG
- Product
- Tibial Component. orthopedic prosthesis. Model Nos: Small, W 16-2817/02 Medium, W 16-2817/05 Large, W 16-2817/07
- Product code
- JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer
- Reason for recall
- There is a risk that blind screws of the modular tibial component cannot be loosened intraoperatively. This may lead to prolongation of surgery due to an intraoperative change in procedure.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- On May 24, 2022, the firm notified affected customers via Urgent Field Safety Notice email. Customers were provided a modified surgical technique providing information for the cementing technique for Tilastan proximal spacers to minimize the risk. No product return is required.
- Quantity in commerce
- 3785
- Distribution
- Domestic distribution to AL, CA, FL, GA, IL, IN, KS, MI, MN, NV, NY, OH, TN, TX, WA. Foreign distribution to Algeria Argentina Australia Austria Belgium Bulgaria Canada China Colombia Denmark Ecuador Estonia Finland France Germany Great Britain Greece Hungary India Indonesia Israel Italy Latvia Libya Lithuania Mexico Netherlands Norway Pakistan Peru Philippines Poland Romania Saudi Arabia Slovakia Slovenia Spain Sudan Sweden Switzerland Thailand Turkey Ukraine United Arab Emirates Uruguy Vietnam
- Product codes and lots
- UDI-DI: 04026575359202 Small, W; 04026575359219 Medium, W; 04026575359226 Large, W; Serial/Lot Numbers: 201002/1683 211025/1543 210202/0275 210302/0308 210314/1872 201002/1687
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1377-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1378-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1379-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1380-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.