Z-1408-2022Class II

Nonsterile convenience kit: regard Item #830014016, LD00199P - C Section Mother

ROi CPS LLC / initiated 2022-06-03 / Terminated / root cause: packaging

ROi CPS LLC began a recall of Nonsterile convenience kit: regard Item #830014016, LD00199P - C Section Mother on . FDA classified it as Class II and gives the reason as "Some nonsterile components that were supposed to be sterile were packaged in nonsterile kits". As of , the recall status on record is Terminated, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1408-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
ROi CPS LLC
Product
Nonsterile convenience kit: regard Item #830014016, LD00199P - C Section Mother.
Product code
OHM Cesarean Section Tray
Reason for recall
Some nonsterile components that were supposed to be sterile were packaged in nonsterile kits.
Root cause tag
packaging
FDA root cause
Process change control
Action
The recalling firm initiated the recall via telephone on 6/3/2022 to the direct consignee of Item 830019017, explained the reason for recall, and instructed the product be quarantined. The recalling firm also contacted the end user who had received distribution from the direct consignee. The end user was informed of the error and was requested to remove the affected lot from their inventory. The recalling firm worked with the end user to have the kits returned to the direct consignee where they were placed in quarantine. Regarding the recall of Item #830099008 and Item #830014016, the direct consignees were notified on 6/6/2022 via telephone requesting the affected kits be placed on hold and moved to a quarantined area. A follow-up email was sent to the recall coordinator for the direct consignees (which was the same recall coordinator) on 6/16/2022 which contained a correction notice. The email explained the issue and said the recalling firm has already performed rework on the kits in the inventory at their location (the two direct consignee locations) so no additional actions were required except for the completion of the reply form to indicate receipt of the notice. The attached Medical Device Correction Notice was dated 6/16/2022 and explained the issue, the affected product, the health risk assessment, and the actions taken by the recalling firm, which included rework of the products that had been returned to the direct consignees. The enclosed replay form was to be completed acknowledging the consignee has read and understands the notice.
Quantity in commerce
40 kits
Distribution
Distribution was made to MO. There was no foreign/military/government distribution.
Product codes and lots
Lot #92605, exp. 8/31/2023, Item #830014016, UDI #10194717110156.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1408-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1407-2022Class IITerminatedVoluntary: Firm initiated
Z-1408-2022Class IITerminatedVoluntary: Firm initiated