Z-1403-2022Class II
RayStation 9B SP1. For radiation treatment planning
RaySearch Laboratories AB (publ) / initiated 2022-06-01 / Terminated / root cause: software
RAYSEARCH LABORATORIES AB began a recall of RayStation 9B SP1. For radiation treatment planning on . FDA classified it as Class II and gives the reason as "An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1403-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation 9B SP1. For radiation treatment planning.
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
- Root cause tag
- software
- FDA root cause
- Software change control
- Action
- The firm contacted all customers via email on June 1, 2022. Customers were provided with a list of workflows where safety related information may not be propagated. Customers were instructed to be aware that treatment course information in RayCare may not always be up to date. Customers should verify critical information in RayStation or RayTreat before treatment course information in RayCare is used for critical decisions. Customers were asked to identify all installed units with the affected software version numbers. The issue will be resolved in mandatory software updates for RayStation, scheduled for market release in June 2022.
- Quantity in commerce
- 1
- Distribution
- Worldwide distribution - US Nationwide distribution in the state of TN and the countries of Belgium and Switzerland.
- Product codes and lots
- UDI-DI: 0735000201029720200310 Build number 9.2.0.483
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1403-2022 | Class II | Terminated | Voluntary: Firm initiated |