Z-1403-2022Class II

RayStation 9B SP1. For radiation treatment planning

RaySearch Laboratories AB (publ) / initiated 2022-06-01 / Terminated / root cause: software

RAYSEARCH LABORATORIES AB began a recall of RayStation 9B SP1. For radiation treatment planning on . FDA classified it as Class II and gives the reason as "An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare". As of , the recall status on record is Terminated, and MDM tags the root cause as software.

Recall record

Recall number
Z-1403-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
RaySearch Laboratories AB (publ)
Product
RayStation 9B SP1. For radiation treatment planning.
Product code
MUJ System, Planning, Radiation Therapy Treatment
Reason for recall
An issue with propagation of treatment course information from RayStation with RayTreat to RayCare has been identified. During some workflows, information may not be propagated to RayCare.
Root cause tag
software
FDA root cause
Software change control
Action
The firm contacted all customers via email on June 1, 2022. Customers were provided with a list of workflows where safety related information may not be propagated. Customers were instructed to be aware that treatment course information in RayCare may not always be up to date. Customers should verify critical information in RayStation or RayTreat before treatment course information in RayCare is used for critical decisions. Customers were asked to identify all installed units with the affected software version numbers. The issue will be resolved in mandatory software updates for RayStation, scheduled for market release in June 2022.
Quantity in commerce
1
Distribution
Worldwide distribution - US Nationwide distribution in the state of TN and the countries of Belgium and Switzerland.
Product codes and lots
UDI-DI: 0735000201029720200310 Build number 9.2.0.483

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1403-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1403-2022Class IITerminatedVoluntary: Firm initiated