Z-1428-2022Class II

ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R

Aesculap Inc / initiated 2020-07-13 / Terminated / root cause: design

Aesculap Implant Systems LLC began a recall of ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R on . FDA classified it as Class II and gives the reason as "Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system". As of , the recall status on record is Terminated, and MDM tags the root cause as design.

Recall record

Recall number
Z-1428-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
Aesculap Inc
Product
ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system.
Root cause tag
design
FDA root cause
Component design/selection
Action
An Urgent Field Safety Notification was issued July 13, 2020 sent USPS certified mail to distributors. The notification identified the product as Ennovate Removal Key (SZ380R) and addressed correct usage of said device, referencing section K of the Surgical Technique. Distributors are to identify and notify customers if further distributed. Distributors and customers are to review the notice, notify end users, and return the "Medical Device Field Safety Notification Form" to Aesculap via email at [email protected].
Quantity in commerce
16
Distribution
United States Nationwide distribution in the states of CA, CO, IL, MI, OH & TX.
Product codes and lots
UDI-DI: 04046964719622

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1428-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1428-2022Class IITerminatedVoluntary: Firm initiated