Z-1426-2022Class II

Sensis Vibe Combo with software version VD12A. A diagnostic and administrative tool supporting hemodynamic...

Siemens Medical Solutions USA, Inc. / initiated 2022-06-15 / Open, Classified / root cause: design

Siemens Medical Solutions USA, Inc began a recall of Sensis Vibe Combo with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies... on . FDA classified it as Class II and gives the reason as "Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1426-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Siemens Medical Solutions USA, Inc.
Product
Sensis Vibe Combo with software version VD12A. A diagnostic and administrative tool supporting hemodynamic catheterizations and/or electrophysiology studies, for cardiac as well as interventional Radiology Model: 11007642
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Sensis/Sensis Vibe SoftwareVD12A, the ComboBox may encounter a disconnect of its communication during the first patient examination of the day or after a longer period of inactivity to the Sensis Vibe system resulting in no vital signs being available.
Root cause tag
design
FDA root cause
Device Design
Action
Siemens issued a Customer Safety Advisory Notice (AX016/22/S) dated 6/15/22 distributed to customers. Letter states reason for recall, health risk and action to take. The BIOS configuration of the ComboBox in the affected systems will be updated to correct the issue. This corrective action will be distributed via Update Instruction AX015/22/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Urgent Safety Advisory Notice. By signing this document, you are acknowledging that you have read and understand the content therein
Quantity in commerce
13 units
Distribution
US Nationwide Distribution AZ FL KY MI OH PR TX
Product codes and lots
UDI-DI: 04056869010205  Serial Numbers:  123021, 123055, 123056, 123058, 122042, 123004, 122006, 122026, 122029, 123047, 123053, 122011, 121014.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1426-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1425-2022Class IIOngoingVoluntary: Firm initiated
Z-1426-2022Class IIOngoingVoluntary: Firm initiated