Z-1299-2022Class II
RayStation 11B
RaySearch Laboratories AB (publ) / initiated 2022-06-10 / Terminated
RAYSEARCH LABORATORIES AB began a recall of RayStation 11B on . FDA classified it as Class II and gives the reason as "This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might...". As of , the recall status on record is Terminated.
Recall record
- Recall number
- Z-1299-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- RaySearch Laboratories AB (publ)
- Product
- RayStation 11B
- Product code
- MUJ System, Planning, Radiation Therapy Treatment
- Reason for recall
- This notice concerns two issues found related to the display of Linear Energy Transfer (LET) in RayStation 11B including some service packs. First, when using a dose threshold for an evaluation dose, LET display might be misleading. Second, a displayed beam-specific LET distribution can sometimes be out of sync with the selected beam.
- FDA root cause
- Under Investigation by firm
- Action
- The firm notified customers of the recall via email on June 10, 2022. The notice was titled "Field Safety Notice, Medical Device Correction #99834." Customers were given precautions and instructed on a workaround. The issue will be resolved in the next version of RayStation, scheduled for market release in June 2022, though this may be subject to market clearance.
- Quantity in commerce
- 26
- Distribution
- US Nationwide distribution in the states of NJ and TX.
- Product codes and lots
- UDI-DI 0735000201042620211208 Software version numbers 12.0.0.932, 12.1.0.1221, 12.0.3.68
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1299-2022 | Class II | Terminated | Voluntary: Firm initiated |