Z-1508-2022Class II

The Cholestech LDX System combines enzymatic methodology and solidphase technology to measure total...

Alere San Diego, Inc. / initiated 2022-06-24 / Open, Classified / root cause: design

Alere San Diego, Inc. began a recall of The Cholestech LDX System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose... on . FDA classified it as Class II and gives the reason as "Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1508-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Alere San Diego, Inc.
Product
The Cholestech LDX System combines enzymatic methodology and solidphase technology to measure total cholesterol, HDL cholesterol, triglycerides and glucose. Samples used for testing can be whole blood from a fingerstick (collected in a lithium heparin-coated capillary tube) or venipuncture. The sample is applied to an Cholestech LDX cassette. Description/Part: Cholestech LDX Lipid Profile GLU Cassette/10-991, 99021, 14-531; Cholestech LDX TC HDL GLU Cassette/10-990; Cholestech LDX Lipid Profile Cassette/10-989; Cholestech LDX TC HDL Cassette/10-987
Product code
CGA Glucose Oxidase, Glucose
Reason for recall
Cholesterol test cassettes package insert is to be updated with new bilirubin interference information because it was found that samples containing bilirubin, at concentrations greater than 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated), can result in low recovery of Triglyceride (TRG) and High-Density Lipoproteins (HDL).
Root cause tag
design
FDA root cause
Device Design
Action
On 6/24/22, correction notices were mailed and emailed to customers who were asked to take the following actions: 1) Test results from patients with bilirubin levels above 2.0 mg/dL (conjugated) or 1.6 mg/dL (unconjugated) should be verified using another test method. 2) Complete and return the Customer Reply Form. 3) If you have forwarded the products listed above to other laboratories, please inform them of this Product Correction and provide to them a copy of this letter. 4) Retain this letter for your laboratory records. Customers with questions are encouraged to contact Technical Service at 877-308-8289.
Quantity in commerce
436,990
Distribution
US: IL, NV, FL, LA, CA, MA, MN, NC, AK, NE, TX, PA, RI, SC, TN, OK, KY, GA, NY, MO, CO, AZ, OH, HI, VA, KS, IN, AL. OUS: GB, ZA, NO, NL, FI, IE, IQ, IT, CH, DE, ES, HU, SI, NA, PT, IS, NG, PL, CY, TN, BH, ID, MY, AU, SG, JP, IN, BD, KR, HK, BR, CO, AR, CA
Product codes and lots
All lots of Part/UDI-DI: 10-991/00893038002654, 99021/00899722002344, 14-531/04571226470831, 10-990/00893038002647, 10-989/00893038002630, 10-987/00893038002616

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1508-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1508-2022Class IIOngoingVoluntary: Firm initiated