Z-1467-2022Class II

TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine...

Medtronic, Inc. / initiated 2022-06-07 / Open, Classified / root cause: design

Covidien Llc began a recall of TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808 on . FDA classified it as Class II and gives the reason as "Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-1467-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic, Inc.
Product
TruClear Hysteroscopic Tissue Removal System Control Unit- designed to meet the requirements of intrauterine tissue removal. Model Number: 7209808
Product code
HIH Hysteroscope (And Accessories)
Reason for recall
Electrical component failure within the control unit. The observed failure mode is a loss of function of the device when used with handpiece and foot switch.
Root cause tag
design
FDA root cause
Device Design
Action
Medtronic issued Urgent Medical Device Recall letter on 6/7/22 (Field Action #: FA1259) via UPS 2-day delivery. For outside U.S. countries, notifications are handled via a regionally approved method. Letter states reason for recall, health risk and action to take: "Identify and quarantine all affected product. The enclosed Customer Notification DetailReport includes affected product for your facility. " Return all affected product in your inventory to Medtronic. a. Product purchased directly from Medtronic: [email protected] b. Product purchased through distributor: [email protected] " Your local Medtronic Representative can assist you as necessary in initiating the return of this product. " Complete the enclosed Customer Confirmation Form and email it to [email protected] " If you have questions or concerns regarding this recall, please contact Medtronic CustomerService at 800-854-3570.
Quantity in commerce
33 units
Distribution
AZ CA CO IL KS MA MI MN MO NY OK PA TN TX TX WA Foreign: United Kingdom Ireland Netherlands Reunion Spain
Product codes and lots
UDI-Device Identifier (GTIN/UPN) 10884521749931 Serial Number: YF1120, YF1126, YF1127, YF1128, YF1129, YF1130, YF1131, YF1133, YF1134, YF1135, YF1137, YF1138, YF1139, YF1140, YF1141, YF1142, YF1150, YF1152, YF1154, YF1156, YF1157, YF1158, YF1161, YF1163, YF1166, YF1167, YF1168, YF1169, YF1172, YF1175, YF1176, YF1177, YF1182

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1467-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1467-2022Class IIOngoingVoluntary: Firm initiated