Z-1429-2022Class II
ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R
Aesculap Inc / initiated 2020-11-18 / Terminated / root cause: design
Aesculap Implant Systems LLC began a recall of ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R on . FDA classified it as Class II and gives the reason as "Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system. Redesign of removal key". As of , the recall status on record is Terminated, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1429-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Aesculap Inc
- Product
- ENNOVATE MIS REMOVALKEY SHORT, product code SZ380R
- Product code
- LXH Orthopedic Manual Surgical Instrument
- Reason for recall
- Fracture or breakage of the spinal fixation arm at the downtube instrument if the 90 degree angle is not met when using the removal key with the spinal fixation system. Redesign of removal key.
- Root cause tag
- design
- FDA root cause
- Component design/selection
- Action
- An Urgent Field Safety Notification was issued November 18, 2020 sent USPS certified mail to distributors regarding removal of Ennovate Removal Key Short (SZ380) from the field. The affected Ennovate MIS Removal Key Short (SZ380R) in the field is to be exchanged with an improved design version under the same article number manufactured after June 01, 2020. Distributors are to identify and notify customers of this action. Distributors and customers are to identify the affected version in inventory and return it to Aesculap and submit the Acknowledgment Form via email at [email protected].
- Quantity in commerce
- 17
- Distribution
- United States Nationwide distribution in the states of CA, CO, MI, NH, OH, SC & TX.
- Product codes and lots
- UDI-DI: 04046964719622
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1429-2022 | Class II | Terminated | Voluntary: Firm initiated |