Z-1495-2022Class II

Merlin PCS 3650 programmer Model 3330 software

Abbott Medical / initiated 2022-06-17 / Open, Classified / root cause: software

St. Jude Medical, Cardiac Rhythm Management Division began a recall of Merlin PCS 3650 programmer Model 3330 software on . FDA classified it as Class II and gives the reason as "There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1495-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Abbott Medical
Product
Merlin PCS 3650 programmer Model 3330 software
Product code
DXY Implantable Pacemaker Pulse-Generator
Reason for recall
There is a potential for programmer software (Merlin PCS) and remote monitoring software application (Merlin.net) to display overestimated predicted battery longevity for certain pacemakers.
Root cause tag
software
FDA root cause
Software design
Action
On June 16, 2022 Abbott issued a "Medical Device Correction" via: XXXX to affected consignees. The customer notification ask consignees to take the following action: 1. Prophylactic device replacement is not recommended, as device functionality, actual longevity, and ERI indicator are not impacted (device functionality remains normal and within specifications). 2. Routine follow-up should remain as per local standard of care and clinical protocol, and ERI should continue to serve as an indicator of the need for device replacement scheduling. 3. Please direct any questions about device longevity to Abbott Technical Support at 1-800-722-3774 (U.S.). 4.Upon programmer software / remote monitoring software update, the improved longevity estimate will be displayed at the patient s next interrogation. Please note that until programmers are updated, a difference in longevity estimates between programmers and remote monitoring (Merlin.net) may be observed.
Quantity in commerce
29,996 programs
Distribution
US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VI VT WA WI WV WY , Guam OUS: ALBANIA ALGERIA ANGOLA Argentina ARMENIA Australia AUSTRIA AZERBAIJAN BAHRAIN Bangladesh BELGIUM BENIN BOLIVIA BOSNIA-HERZ. BRAZIL BULGARIA Burnei Cambodia CAMEROON Canada Chile China Colombia COTE D'IVOIRE CUBA CYPRUS CZECH REPUBLIC DENMARK DOMINICAN REP. ECUADOR EGYPT EL SALVADOR ESTONIA ETHIOPIA FAROE ISLANDS FINLAND FRANCE GEORGIA GERMANY GREECE GREENLAND GUATEMALA Hong Kong HUNGARY ICELAND India Indonesia IRELAND ISRAEL ITALY JORDAN KAZAKHSTAN KOSOVO KUWAIT KYRGYZSTAN LATVIA LEBANON LIBYA LIECHTENSTEIN LITHUANIA LUXEMBOURG MACEDONIA Malaysia MALTA MAURITIUS Mexico MONGOLIA MONTENEGRO MOROCCO Myanmar Nepal NETHERLANDS New Zealand NORWAY OMAN Pakistan PALESTINE PANAMA PARAGUAY PERU Philippines POLAND PORTUGAL QATAR ROMANIA RUSSIAN FED. SAUDI ARABIA SENEGAL Singapore SLOVAKIA SLOVENIA SOUTH AFRICA South Korea SPAIN Srilanka SUDAN SWEDEN SWITZERLAND Taiwan Thailand TRINIDAD TUNISIA TURKEY UKRAINE United Arab Emirates UNITED KINGDOM UZBEKISTAN VENEZUELA Vietnam VIRGIN ISLAND YEMEN ZAMBIA
Product codes and lots
Software Model 3330 v26.0.1 rev 1 or earlier; UDI/DI 05414734509725

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1495-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1495-2022Class IIOngoingVoluntary: Firm initiated
Z-1496-2022Class IIOngoingVoluntary: Firm initiated
Z-1497-2022Class IIOngoingVoluntary: Firm initiated