Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures
Stryker Instruments began a recall of Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures on . FDA classified it as Class II and gives the reason as "There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1478-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Corp.
- Product
- Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.
- Product code
- HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
- Reason for recall
- There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.
- FDA root cause
- Process change control
- Action
- On June 24, 2022, the firm notified customers via Urgent Medical Device Recall Letters. Customers were instructed to locate and quarantine any affected products found in their inventory. Upon receiving the customer's completed response form, the firm will provide a shipping label to return any recalled product. Upon receipt of the recalled product, Stryker will contact the customer to arrange for replacement product.
- Quantity in commerce
- 543
- Distribution
- Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.
- Product codes and lots
- UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1478-2022 | Class II | Ongoing | Voluntary: Firm initiated |