Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures

Stryker Corp. / initiated 2022-06-24 / Open, Classified

Stryker Instruments began a recall of Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures on . FDA classified it as Class II and gives the reason as "There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1478-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Stryker Corp.
Product
Stryker 2.3mm Tapered Router, Catalog no. 5820-071-023 instrument used in craniotomy procedures.
Product code
HBE Drills, Burrs, Trephines & Accessories (Simple, Powered)
Reason for recall
There is a potential for the core diameter to be undersized, which may lead to the router breaking. A break can lead to unintended metal fragments in the surgical site, resulting in the potential for surgical intervention or damage to critical neurological or vascular structures.
FDA root cause
Process change control
Action
On June 24, 2022, the firm notified customers via Urgent Medical Device Recall Letters. Customers were instructed to locate and quarantine any affected products found in their inventory. Upon receiving the customer's completed response form, the firm will provide a shipping label to return any recalled product. Upon receipt of the recalled product, Stryker will contact the customer to arrange for replacement product.
Quantity in commerce
543
Distribution
Worldwide distribution - US Nationwide distribution in the states of Arizona, California, Florida, Georgia, Iowa, Michigan, Missouri, New Mexico, North Carolina, Ohio, Texas, Virginia and the countries of Canada, Netherlands, and Thailand.
Product codes and lots
UDI-DI (GTIN): 07613327294910 Affected Lots: 22011017 21330017 20216017 20140017 20139027

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1478-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1478-2022Class IIOngoingVoluntary: Firm initiated