ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in...
DePuy Orthopaedics, Inc. began a recall of ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525 on . FDA classified it as Class II and gives the reason as "Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1419-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Depuy Orthopaedics Inc.
- Product
- ATTUNE Measured Sizing and Rotation Guide-to size the femur and align rotation of the bone cuts in preparation for the femoral implant Part Number: 254400525
- Product code
- HWT Template
- Reason for recall
- Manufactured with a Galvanized High Carbon Steel spring instead of the correct 302 Stainless steel spring. has the potential to oxidize and corrode in the ATTUNE Measured Sizing & Rotation Guide, resulting in surgical delay, debris in surgical field, Osteolysis, Poor Joint Mechanics, Soft Tissue Irritation, Pain
- Root cause tag
- manufacturing process
- FDA root cause
- Process control
- Action
- DePuy Synthes issued Urgent Medical Device Correction Letter on June 29 2022. Letter states reason for recall, health risk and action to take: 1. Examine your inventory immediately for the subject lots and quarantine the products. 2. To minimize disruption, replacement of the spring will be completed by a Field Service Representative. You will be contacted directly by a member of the Field Service Team. 3. Familiarize yourself with the content of this letter and forward this notice to any personnel in your facility who need to be informed. 4. Complete all fields, sign, and return the attached business response form (BRF) on the last page of this letter to [email protected] within three (3) business days of receipt of this medical device correction letter. Please include FA 2126657 - Attune Measured Sizing & Rotation Guide in the email subject line. 5. Complete the BRF even if you do not have any of the subject product in your current inventory. 6. If any of the subject product has been forwarded to another facility, contact that facility, and provide a copy of this notice to the relevant personnel. 7. Post a copy of this notice in a visible area for awareness of this correction. Contact your local DePuy Synthes Sales Consultant. For Medical Information Requests, please visit our website: https://www.jnjmedicaldevices.com/mir
- Quantity in commerce
- 872 units
- Distribution
- Nationwide Foreign: Argentina, Australia, Austria, Bahrain, Belgium, Brazil, Canada, Chile, China, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hong Kong, India, Ireland, Israel, Italy, Japan, Kuwait, Lithuania, Mauritius, Mexico, Morocco, Netherlands, New Zealand, Norway, Peru, Poland, Portugal, Puerto Rico, Romania, Saudi Arabia, Serbia, Singapore, South Africa, South Korea, Spain, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom, Vietnam
- Product codes and lots
- GTIN: 10603295434061 Lots: AB4536235, AB4733296, AB4841493, AB4810808, AB4841495, AB4904565, AB4967214, AB5004557, and AB5034303
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1419-2022 | Class II | Ongoing | Voluntary: Firm initiated |