Z-1700-2022Class II
Various neurology packs
American Contract Systems, Inc. / initiated 2022-06-16 / Open, Classified / root cause: sterility
American Contract Systems, Inc. began a recall of Various neurology packs on . FDA classified it as Class II and gives the reason as "Product was exposed to multiple sterilization cycles without validation for multiple exposures". As of , the recall status on record is Open, Classified, and MDM tags the root cause as sterility.
Recall record
- Recall number
- Z-1700-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- American Contract Systems, Inc.
- Product
- Various neurology packs
- Product code
- OJG Neurological Tray
- Reason for recall
- Product was exposed to multiple sterilization cycles without validation for multiple exposures.
- Root cause tag
- sterility
- FDA root cause
- Process change control
- Action
- The recalling firm began notifying customers of the recall beginning June 13, 2022. Customer notification was initially made via email and an "Urgent Medical Device Recall" letter dated June 16, 2022. Customers are asked to review inventory and segregate affected product, which should be identified with the Bag Serial Number. Affected product should be destroyed, with evidence of destruction provided back to the recalling firm. A Field Action Response Form should be completed and returned to [email protected] as soon as possible, even if the customer has no inventory. If you have any questions, please call 816-920-5846.
- Quantity in commerce
- 37 units
- Distribution
- Distribution within the United States to health systems and distributors in AZ, FL, IA, IL, MA, MN, MO, OH, NE, PA, RI, SD, and TX
- Product codes and lots
- Tray Number, Sterilization Lot, Lot Number, Bag Serial Number 1) LSRLMNEINTC, 2105181, 870211, 63990989 2) LSRSWRSPNC, 2105251, 861211, 63990357 3) LSRLMNEUROG, 2106161, 845211, 44139500 4) LSRSMHNEURE, 2108101, 793211, 44139500 5) LSRNEURCRANIH, 2108171, 776211, 64123199 6) LSRLMNEUROG, 2110011, 730211, 64316283 7) UINR44M, 2104162, 896211, 64033036 8) TNNE17Y, 2109081, 750211, 64224067 9) SLCR80E, 2104262, 884211, 63986071-006 10) GISP04A, 2104293, 881211, 63985242-006 11) FHNE25Y, 2107083, 811211, 64110740-006 12) BHLA24L, 2108253, 763211, 64205048-006 13) CBAC47F, 2109092, 749211, 64209378-006 14) SLCN72B, 2109161, 741211, 64208359-006 15) BHCR54D, 2104132, 901211, 64072795-006 16) BHCR54D, 2104132, 901211, 64072796-006 17) BHCR54D, 2104132, 901211, 64072273-006 18) BHLM30D, 2105111, 874211, 64062457-006 19) BHCR54D, 2105252, 861211, 64063353-006 20) BHLM30D, 2107091, 819211, 64088774-006 21) BHLM30D, 2107091, 819211, 64088775-006 22) BHLM30D, 2107091, 819211, 64088776-006 23) BHCR54D, 2107161, 810211, 64088824-006 24) BHCR54D, 2107161, 810211, 64088823-006 25) BHLM30D, 2108202, 770211, 64001247-006 26) BHLM30D, 2108202, 770211, 64001248-006 27) BHLM30D, 2108202, 770211, 64001249-006 28) BHLM30D, 2108202, 770211, 64001250-006 29) BHCR54D, 2109022, 757211, 63998620-006 30) BHCR54D, 2109022, 757211, 63998619-006 31) BHCR54D, 2109022, 757211, 63998618-006 32) UMNA22B, 2109292, 728211, 64310294-006 33) DVSP33E, 2104302, 884211, 64064626-006 34) MCSP61F, 2107021, 819211, 64128815-006 35) SJSC27F, 2107091, 813211, 44173743-005 36) SANE32G, 2108091, 784211, 44170921-005 37) MCSP61F, 2108091, 785211, 64138850-006
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1689-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1690-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1691-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1692-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1693-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1694-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1695-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1696-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1697-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1698-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1699-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1700-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1701-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1702-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1703-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1704-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1705-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.