Z-1549-2022Class II
Wolf-Pak Premium Dressing Kit with GuardVA Cs/30 Ref: DC5030LF
Trinity Sterile, Inc. / initiated 2022-06-29 / Terminated / root cause: labeling and UDI
Trinity Sterile, Inc. began a recall of Wolf-Pak Premium Dressing Kit with GuardVA Cs/30 Ref: DC5030LF on . FDA classified it as Class II and gives the reason as "Kits mislabeled as latex-free, the packaging for component 94-7002 Non-Adherent Dressing contains natural rubber latex. Potential risk to to those with latex allergy". As of , the recall status on record is Terminated, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1549-2022
- Classification
- Class II
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Trinity Sterile, Inc.
- Product
- Wolf-Pak Premium Dressing Kit with GuardVA Cs/30 Ref: DC5030LF
- Product code
- OXQ Dressing Change Tray
- Reason for recall
- Kits mislabeled as latex-free, the packaging for component 94-7002 Non-Adherent Dressing contains natural rubber latex. Potential risk to to those with latex allergy
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling design
- Action
- Trinity Medical issued Urgent Medical Device Correction Letter dated June 30, 2022 via email as a follow-up to the call on 6/29/22. Letter states reason for recall, health risk and action to take: 1. Inspect your inventory for the affected products and lot numbers identified on the attached spreadsheet. If you have affected stock in inventory, immediately discontinue use and quarantine the products. If you have transferred any of the affected kits to a third party, please communicate this information to them. 2. Notify Trinity. Please complete the enclosed Customer Response Form as soon as possible, indicating whether you understand this medical device correction. Fax it to 410-860-2913 or email it to [email protected]. 3. Relabel the affected product. Once Trinity receives your completed Customer Response Form, Trinity will provide you with brightly colored stickers to attach to each case and kit. If you have any questions, please contact Customer Service at 410-860-5123.
- Quantity in commerce
- 1,513 cases (45,390 eaches)
- Distribution
- US Nationwide distribution in the states of FL, PA.
- Product codes and lots
- UDI-DI: 10850007185760 Case 00850007185763 Kit Lot Numbers: 600000 600001 600002 600003 600004 600005 600006 600007 600008
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1548-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1549-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1550-2022 | Class II | Terminated | Voluntary: Firm initiated | ||
| Z-1551-2022 | Class II | Terminated | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.