Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1562-2022Class II

DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb...

DeRoyal Industries, Inc. / initiated 2022-07-15 / Terminated

DeRoyal Industries Inc began a recall of DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3)... on . FDA classified it as Class II and gives the reason as "The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure". As of , the recall status on record is Terminated.

Recall record

Recall number
Z-1562-2022
Classification
Class II
Status
Terminated
Initiated
Posted
Terminated
Recalling firm
DeRoyal Industries, Inc.
Product
DeRoyal limb holders: (1) DeRoyal Limb Holder, Wrist, Fixed, Quick Release, REF M8124; (2) DeRoyal Limb Holder, Ankle, Fixed, Quick Release, REF M8125; (3) DeRoyal Security Cuffs, Wrist and Ankle, Universal, REF M8136; and (4) DeRoyal Security Cuffs, Heavy Duty Vertical Psychiatric, University, REF M8137.
Product code
FMQ Restraint, Protective
Reason for recall
The webbing strap was laced improperly through the teeth of the buckle allowing the webbing strap to slide and was, therefore, not secure.
FDA root cause
Under Investigation by firm
Action
The recalling firm issued recall letters dated 7/15/2022 via email to the distributors and sale representative, and via overnight mail to the institution on 7/15/2022. The letter with an attached Affected Products Listing explained the reason for recall, the health risk, and provided actions for the customer to take, which included the following: Using the attached Affected Products Listing, identify the affected part numbers and lot numbers in inventory and place them in quarantine pending return. Complete the Notice of Return Form and return it to the recalling firm via FAX or email no later than 8/26/2022. If the product has been further distributed, the customer is requested to contact their customers by either forwarding the recall notification or requesting the recalling firm notify them.
Quantity in commerce
36 eaches and 7 pairs
Distribution
US Nationwide distribution in the states of FL, IN, NV, RI, TN, TX, and WV.
Product codes and lots
(1) M8124 - Lot 57372551 / exp 06/06/2032; Lot 57285721 / exp 05/22/2032; Lot 57369651 / exp 06/06/2032; Lot 57497652 / exp 6/27/2032; GTIN: 00749756780728; (2) M8125 - Lot 5746744 / exp 6/20/2032; Lot 57419195 / exp 6/13/2032; Lot 57420031 / exp 6/14/2032; GTIN: 00749756780735; (3) M8136 - Lot 57501121 / exp 6/27/2032; GTIN: 00749756837828; (4) M8137 - Lot 57462151 / exp 6/20/2032; GTIN: 00749756849234.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1562-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1562-2022Class IITerminatedVoluntary: Firm initiated