Z-1641-2022Class II
T-Spot.tb Ref Tb.300 US
Oxford Immunotec Ltd / initiated 2022-07-07 / Open, Classified / root cause: manufacturing process
OXFORD IMMUNOTEC LTD began a recall of T-SPOT.TB REF TB.300 US on . FDA classified it as Class II and gives the reason as "Due to equipment failure, affected products were stored at temperatures outside the validated range of 2-8 degree C for a period longer than the validated storage time". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-1641-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Oxford Immunotec Ltd
- Product
- T-Spot.tb Ref Tb.300 US
- Product code
- OJN Mycobacterium Tuberculosis, Cell Mediated Immune Response, Enzyme-Linked Immunospot Test
- Reason for recall
- Due to equipment failure, affected products were stored at temperatures outside the validated range of 2-8 degree C for a period longer than the validated storage time.
- Root cause tag
- manufacturing process
- FDA root cause
- Storage
- Action
- On 07/07/2022, Oxford Immunotec Limited send a Field Safety Notice to customers informing them that the firm has become aware that products distributed may have been stored outside of the storage specifications for a short period. Customer are asked to quarantined the affected products until further notice. On 07/13/2022, the firm communicated a follow-up "URGENT ACTION REQUIRED" Letter to customer asking to confirm that affected products have been segregated into a quarantine area and have not been used to produce patient results. On 07/20/2022, the firm sent a "VOLUNTARY PRODUCT RECALL NOTIFICATION" informing customers that Oxford Immunotec Limited is conducting a full recall of the affected products and asking customers to now destroy the affected products following these steps: 1. Ensure Recalled Product remains in Quarantine. As per our previous communication dated July-13-2022, please ensure all stock of the above products remain in a secure location pending destruction. 2. Complete Tracking/Verification Form. Complete and return the enclosed tracking/verification form, following the directions on this page and the Tracking/Verification Form. Your Oxford Immunotec sales representative can assist you in completing the form. 3. Indicate Destroyed Products on Tracking/Verification Form. The tracking/verification form lists the [product name] and product descriptions. Please mark the number of each type of [product] that you are destroying on this form. On or after 08/03/2022, Oxford Immunotec Ltd communicated an updated/follow-up customer notification with included the potential risks/harms/hazards associated with the use of the affected products. For questions or assistance - contact Chris Dark at +44 1235 442780 or Maureen Garner at 855-NWRS-222 (855-697-7222), or alternatively any Oxford Immunotec Sales Representatives.
- Quantity in commerce
- 81 units
- Distribution
- U.S.: AL, CA, DC, GA, IN, OR, and WA O.U.S.: None
- Product codes and lots
- Model/Catalog Number: TB.300 US UDI-DI Code: 15051716000305 Lot Number: TEC4000098
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1641-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1642-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1643-2022 | Class II | Ongoing | Voluntary: Firm initiated |