Z-1649-2022Class II
Metalogix Self-Drill Half-Pin, a component of the Revolution External Fixation System. Orthopedic. Metalogix...
Fusion Orthopedics USA, LLC / initiated 2020-12-28 / Open, Classified / root cause: design
New Standard Device Inc began a recall of Metalogix Self-Drill Half-Pin, a component of the Revolution External Fixation System. Orthopedic. Metalogix Self-Drill Half-Pin 5x30x215mm REF 620030... on . FDA classified it as Class II and gives the reason as "Complaints were received for Self-drill Half-Pin tip breakages occurring during surgery". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1649-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Fusion Orthopedics USA, LLC
- Product
- Metalogix Self-Drill Half-Pin, a component of the Revolution External Fixation System. Orthopedic. Metalogix Self-Drill Half-Pin 5x30x215mm REF 620030 Metalogix Self-Drill Half-Pin 5x40x215mm REF 620040 Metalogix Self-Drill Half-Pin 5x60x215mm REF 620060 Metalogix Self-Drill Half-Pin 5x80x215mm REF 620080
- Product code
- KTT Appliance, Fixation, Nail/Blade/Plate Combination, Multiple Component
- Reason for recall
- Complaints were received for Self-drill Half-Pin tip breakages occurring during surgery.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 12/28/2020, Metalogix send a "Field Notice: Self-Drill Half-Pin" via email. The Notice only informed customer that the Citieffe Half-Pins will temporarily be used to replace the current 5mm Self-Drill Half-Pins. A contact number for Mr. Elton Ely, QA Manager was provided in the notice 833-659-2019 ext. 6
- Quantity in commerce
- 410 total devices
- Distribution
- U.S.: AZ, GA, IL MT, NC, NJ, NV, OH, PA, TX, and VA O.U.S.: None
- Product codes and lots
- Part Number / DESCRIPTION / UDI-DI Code / Lot Numbers: 620030 / Self-Drill Half-Pin 5x30x215mm / 854641008198 / A29418X, A17219CA 620040 / Self-Drill Half-Pin 5x40x215mm / 854641008204 / A29418Y, A17219CB 620060 / Self-Drill Half-Pin 5x60x215mm / 854641008211 / A29418Z, A17219CC 620080 / Self-Drill Half-Pin 5x80x215mm / 854641008228 / A29518A, A17219CD
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1649-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY