Z-1676-2022Class II

VITROS XT 7600 Integrated System with Version 3.7.2 and below

Ortho-Clinical Diagnostics, Inc. / initiated 2022-07-19 / Open, Classified

Ortho-Clinical Diagnostics, Inc. began a recall of VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461 on . FDA classified it as Class II and gives the reason as "Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer...". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1676-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Ortho-Clinical Diagnostics, Inc.
Product
VITROS XT 7600 Integrated System with Version 3.7.2 and below Product Code: 6844461
Product code
JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Reason for recall
Performing a system shutdown on VITROS XT 3400 and VITROS XT 7600 Systems may cause a positive or negative shift in the electrical current supplied to the LED used in conjunction with the Digital Imaging Reflectometer. Potential exists for either a positive or negative shift in results reported by the VITROS System.
FDA root cause
Under Investigation by firm
Action
Ortho CLinical issued Urgent Medical Device Correction letter on 19 July 2022 to consignees who received a VITROS XT3400 or a VITROS XT7600 System notified via (CL2020-176) of the issue and instructed to: " If your VITROS XT 3400 and/or VITROS XT 7600 Systems require a system shutdown, routine Quality Control processing for MicroSlide assays must be performed per your laboratory procedure, prior to processing patient samples, to determine if the anomaly has occurred. " If QC results are outside your established ranges, Ortho recommends performing a second system shutdown and re-processing QC fluids for those assays that were outside your established ranges. If QC results are still outside established ranges following a second system shutdown, contact Ortho Care" Technical Solutions Center for guidance. " Ortho recommends using the same vial of QC fluids (where possible) throughout these steps to determine if the anomaly has occurred. Refer to the control storage and stability documentation for fluid stability guidelines once opened or reconstituted. " Post this notification by each VITROS XT 3400 and/or XT 7600 System until this issue is resolved. Ortho is currently investigating this issue; a follow-up communication will be released once our investigation is complete. " Complete the enclosed Confirmation of Receipt no later than July 26, 2022. If you have further questions, please contact Ortho Care" Technical Solutions Center at 1-800-421-3311.
Quantity in commerce
1148 units
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Bermuda DV04, Brazil, Canada L3R 4G5 , Chile, China, Colombia, Denmark, France, France, Germany, India, Italy, Japan, Mexico, Norway, Portugal, Russia, Singapore 609917, Spain, Sweden, The Netherlands, United Kingdom.
Product codes and lots
UDI-DI: 10758750031610 Systems start with 7600XXXX where XXXX is unique for each analyzer

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1676-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1675-2022Class IIOngoingVoluntary: Firm initiated
Z-1676-2022Class IIOngoingVoluntary: Firm initiated