MEGA 8Fr 50cc Intra-Aortic Balloon Catheter

Datascope Corp. / initiated 2022-08-05 / Open, Classified / root cause: supplier

Datascope Corporation began a recall of MEGA 8Fr 50cc Intra-Aortic Balloon Catheter on . FDA classified it as Class II and gives the reason as "Certain lots containing undersized dilator". As of , the recall status on record is Open, Classified, and MDM tags the root cause as supplier.

Recall record

Recall number
Z-1685-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Datascope Corp.
Product
MEGA 8Fr 50cc Intra-Aortic Balloon Catheter
Product code
DSP System, Balloon, Intra-Aortic And Control
Reason for recall
Certain lots containing undersized dilator.
Root cause tag
supplier
FDA root cause
Vendor change control
Action
The firm distributed an Urgent Medical Device Removal notification, by mail, dated August 5, 2022. The notification informs customers that certain lots of 8Fr. 50cc IAB catheter and insertion kits are being recalled due to an undersized vessel dilator. During a sheathless insertion, the undersized inner diameter may result in the inability to insert the guidewire through the inner lumen of the dilator and the undersized outer diameter may reduce dilation of the vessel. If used in a procedure, recalled product could result in a procedural delay, gas emoli, and vessel injury. Customers are asked to examine their inventory to determine if they have nay recalled product and to quarantine any affected units. Customers with recalled product are to contact Datascope Corp./Getinge Customer Service at 888-9GETUSA (888-943-8872, option 2) between the hours of 8 AM and 6 PM to request a return authorization number to receive full credit. Distributors who have shipped any product are to forward the notification to customers. The recall notification is to be forwarded to all users at hospitals/facilities. Customers are asked to complete and return the recall response form by email to [email protected] or by fax to 877-647-1921.
Quantity in commerce
10,427 units
Distribution
Awaiting domestic distribution. Foreign: Australia, Belgium, Canada, Czech Republic, France, Germany, Hong Kong, Iceland, Ireland, Italy, Netherlands, New Zealand, Norway, Poland, Portugal, Saudi Arabia, Slovenia, South Africa, Spain, Sweden, Switzerland, United Arab Emirates, United Kingdom.
Product codes and lots
Code information reads "UDI-DI, Lot No., Finished Goods Part No.": 10607567109619, 3000184437, 0684-00-0296-01U; 10607567107301, 3000171501, 0684-00-0498-01; 10607567107301, 3000183312, 0684-00-0498-01; 10607567107301, 3000183313, 0684-00-0498-01; 10607567107301, 3000186654, 0684-00-0498-01; 10607567107301, 3000186655, 0684-00-0498-01.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1685-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1685-2022Class IIOngoingVoluntary: Firm initiated
Z-1686-2022Class IIOngoingVoluntary: Firm initiated
Z-1687-2022Class IIOngoingVoluntary: Firm initiated