Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta...
Medtronic Neuromodulation began a recall of Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta implantable neurostimulators for pain therapy on . FDA classified it as Class II and gives the reason as "Potential for Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 to display an Unexpected Device Error Code 1502 Message, and the user will be unable to perform programming of the Vanta Implantable...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-1571-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic Neuromodulation
- Product
- Vanta Clinician Programmer Application, Model A71200. For programming of the Medtronic Model 977006 Vanta implantable neurostimulators for pain therapy.
- Product code
- LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief
- Reason for recall
- Potential for Vanta Clinician Programmer Application (CP App) A71200 v2.0.2455 to display an Unexpected Device Error Code 1502 Message, and the user will be unable to perform programming of the Vanta Implantable Neurostimulator (INS) Model 977006.
- FDA root cause
- Under Investigation by firm
- Action
- On about 07/26/2022, Medtronic notified consignees via mail courier service with a letter titled "URGENT: MEDICAL DEVICE CORRECTION." The letter instructed consignees to perform an initial interrogation of the Vanta INS and tap start usage, tap implant device workflow, click start, and on the next screen exit the session, and to initiate a second interrogation and if there is no "Unexpected Device Error Code 1502" code the INS maybe used. If the INS cannot be interrogated, contact Medtronic Technical Services at 1-800-707-0933. Additionally, consignees were instructed to complete and return the customer confirmation form provided.
- Quantity in commerce
- 2,920 downloaded software applications
- Distribution
- Worldwide distribution - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV, and Puerto Rico. The countries of Austria, Belgium, Canada, Canary Islands, Cayman Islands, Czech Republic, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Luxembourg, Netherlands, Norway, Poland, Portugal, San Marino, Slovakia, Slovenia, Spain, Sweden, Switzerland, United Kingdom.
- Product codes and lots
- Model A71200; Software Version 2.0.2455