BD MaxGuard Extension Set with 2 Needleless Y-Site, REF: MX9059
CAREFUSION began a recall of BD MaxGuard Extension Set with 2 Needleless Y-Site, REF: MX9059 on . FDA classified it as Class II and gives the reason as "Administration set, and extension set product labels list expiration year 2027 but the labels should list expiration year 2025. If expired product is used, the integrity of the product may be compromised, and may result...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-1757-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- CareFusion 303, Inc
- Product
- BD MaxGuard Extension Set with 2 Needleless Y-Site, REF: MX9059
- Product code
- FPA Set, Administration, Intravascular
- Reason for recall
- Administration set, and extension set product labels list expiration year 2027 but the labels should list expiration year 2025. If expired product is used, the integrity of the product may be compromised, and may result in potential leaks and risk of infection to the patient and healthcare worker.
- Root cause tag
- labeling and UDI
- FDA root cause
- Error in labeling
- Action
- On 8/1/22 Correction-recall notices were emailed or mailed to customers and distributors who were asked to do the following: 1) Destroy affected devices following your institutions process for destruction. 2) Share this notification with all users within your facility network to ensure all are aware. The correction-recall notice should also be shared with any organization where the potentially affected product has been transferred. 3) Complete and return the response form. In addition distributors were informed that the recalling firm plans to notify all affected customers directly so they were asked to provide a a list of customers who received affected devices. The distributor response form has a check box if the distributor declines to provide a customer list to the recalling firm because they have notified customers about the correction-recall. A copy of the customer communication was provided to the distributor. If customers or distributors require additional assistance they are aske to call 1-844-823-5433, Mon-Fri 8:00 AM - 5:00 PM CT or [email protected]
- Quantity in commerce
- 17,250
- Distribution
- US: TN, FL, LA, VA, MI, OH, NC, WA, UT, TX, NV, AL, AR, IA, MA, CO, NY, SC, IN, KY, MO, MD, PA, CA, IL, SD, GA, MS, AZ, WV, ID, WY, VT, NM, OR, NJ, OK, KS, MN
- Product codes and lots
- UDI-DI: 10885403237584, Lots: 22029431, 22029432
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1752-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1753-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1754-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1755-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1756-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1757-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1758-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1759-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1760-2022 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.