Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1714-2022Class II

Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-80 ORDER NUMBER (GPN)...

Cook Incorporated / initiated 2022-07-19 / Open, Classified / root cause: packaging

Cook Incorporated began a recall of Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-80 ORDER NUMBER (GPN): G56172. For vascular catheter positioning and exchange... on . FDA classified it as Class II and gives the reason as "Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-1714-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Cook Incorporated
Product
Roadrunner UniGlide Hydrophilic Wire Guide REFERENCE PART NUMBER (RPN): HPWA-35-80 ORDER NUMBER (GPN): G56172. For vascular catheter positioning and exchange in diagnostic and interventional procedures.
Product code
DQX Wire, Guide, Catheter
Reason for recall
Product did not meet the acceptance criteria for packaging testing, which included 3-year age acceleration testing, therefore, the sterility of the device may be compromised.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
Cook notified consignees by Urgent Medical Device Recall Letter on July 19, 2022. Letter states reason for recall, health risk and action to take: Examine inventory immediately to determine if you have affected product(s) and quarantine any affected product that remains unused. Immediately cease all further distribution and use of the affected products. 2. Return the affected product(s) to Cook Medical with a copy of the Acknowledgement and Receipt Form to receive a product credit. Refer to the Acknowledgement and Receipt Form for return instructions. 3. Please complete the Acknowledgement and Receipt Form within 5 business days of receiving this letter regardless of whether you have affected product(s) on hand. You may return the Form via fax (812.339.7316) or email ([email protected]). 4. This notice must be shared with appropriate personnel, including down to the user level, within your organization or with any organization where the potentially affected devices have been transferred. 5. Please immediately report adverse events to Cook Medical Customer Relations by phone at 800.457.4500 or 812.339.2235, Monday through Friday between 7:30am and 5:00pm (Eastern Time), or by email to: [email protected].
Quantity in commerce
75 units
Distribution
Nationwide Foreign: AR AT AU BE BG CA CA CH CL CN CZ DE DK ES FR GB GR GU HU IL IT JO JP KR KW MX MY NC NL PE PL PR PT PY RU SE SI TR UY ZA
Product codes and lots
UDI-DI: (01)00827002561724- Lot Numbers: 14156569

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1714-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1706-2022Class IIOngoingVoluntary: Firm initiated
Z-1707-2022Class IIOngoingVoluntary: Firm initiated
Z-1708-2022Class IIOngoingVoluntary: Firm initiated
Z-1709-2022Class IIOngoingVoluntary: Firm initiated
Z-1710-2022Class IIOngoingVoluntary: Firm initiated
Z-1711-2022Class IIOngoingVoluntary: Firm initiated
Z-1712-2022Class IIOngoingVoluntary: Firm initiated
Z-1713-2022Class IIOngoingVoluntary: Firm initiated
Z-1714-2022Class IIOngoingVoluntary: Firm initiated
Z-1715-2022Class IIOngoingVoluntary: Firm initiated
Z-1716-2022Class IIOngoingVoluntary: Firm initiated
Z-1717-2022Class IIOngoingVoluntary: Firm initiated
Z-1718-2022Class IIOngoingVoluntary: Firm initiated