Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1520-2022Class I

NIM EMG Endotracheal Tube, Part Numbers: a) REF 8229308 b) REF 8229307 c) REF 8229306

Medtronic Xomed, Inc. / initiated 2022-04-29 / Open, Classified

Medtronic Xomed, Inc. began a recall of NIM EMG Endotracheal Tube, Part Numbers: a) REF 8229308 b) REF 8229307 c) REF 8229306 on . FDA classified it as Class I and gives the reason as "The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1520-2022
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic Xomed, Inc.
Product
NIM EMG Endotracheal Tube, Part Numbers: a) REF 8229308 b) REF 8229307 c) REF 8229306
Product code
ETN Stimulator, Nerve
Reason for recall
The firm received reports of events related to airway obstruction while using NIM Standard Reinforced EMG Endotracheal Tubes & NIM CONTACT Reinforced EMG Endotracheal Tubes.
FDA root cause
Under Investigation by firm
Action
Medtronic sent an URGENT: MEDICAL DEVICE SAFETY NOTICE to impacted customers on 29-APR-2022 by mail. The letter explained the problem and provided Recommended Actions when using NIM Standard Reinforced EMG Endotracheal Tube & NIM CONTACT" Reinforced EMG Endotracheal Tube and airway obstruction is encountered. Medtronic posted this issue to its US Product Performance and Advisories Information for Healthcare Professionals website on 3-May-2022. If you have any questions, contact your Medtronic ENT representative. Within the US, beginning 30-Jan-2024, Medtronic distributed a follow-up a notice via mail courier to all impacted consignees about availability of the updated IFU. On about 07/08/2024, Medtronic distributed an additional notification to consignees to notify them to remove all lots of the NIM Contact" EMG Reinforced Endotracheal Tubes and NIM" (Standard) EMG Reinforced Endotracheal Tubes. Customers were instructed to not use the affected devices and to identify, segregate, quarantine, and return affected products to Medtronic using the Customer Confirmation Form. https://www.medtronic.com/content/dam/medtronic-wide/public/united-states/products/ear-nose-throat/nim-contact-emg-reinforced-endotracheal-tubes.pdf.
Quantity in commerce
1,133,612 units
Distribution
Worldwide distribution: US (nationwide) and OUS (foreign) countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Finland, France, French Guiana, Germany, Greece, Hong Kong, Hungary, Iceland, India, Iran, Islamic Republic Of, Ireland, Israel, Italy, Japan, Korea, Republic Of, Kosovo, Kuwait, Libya, Lithuania, Luxembourg, Mexico, Netherlands, New Zealand, North Macedonia, Northern Mariana Islands, Norway, Oman, Panama, Poland, Portugal, Puerto Rico, Qatar, Reunion, Romania, Russian Federation, Saudi Arabia, Slovakia, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Turkey, United Arab Emirates, United Kingdom, and Viet Nam
Product codes and lots
a) REF 8229308; UDI 00643169789548, 00763000745837, 00763000882402; ALL LOTS. b) REF 8229307; UDI 00643169789531, 00763000745820, 00763000882396; ALL LOTS. c) REF 8229306; UDI, 00643169789524, 00763000745813, 00763000882389; ALL LOTS.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1520-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1519-2022Class IOngoingVoluntary: Firm initiated
Z-1520-2022Class IOngoingVoluntary: Firm initiated
Z-1521-2022Class IOngoingVoluntary: Firm initiated