Z-1563-2022Class II

DigitalDiagnost C50 1.1

Philips North America LLC / initiated 2022-02-25 / Open, Classified / root cause: software

Philips North America Llc began a recall of DigitalDiagnost C50 1.1 on . FDA classified it as Class II and gives the reason as "Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-1563-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Philips North America LLC
Product
DigitalDiagnost C50 1.1
Product code
KPR System, X-Ray, Stationary
Reason for recall
Following a system restart where the Table Height 2 (TH2) is the default setting and then switched to Wallstand VS2, operators will see an incorrect orientation of the image on the first examination due to an issue in the firmware of the Wallstand VS2 board. The system will rotate the amplimat field selection by 90 degrees. The wrong amplimat field selection may cause an incorrect dose of radiation to occur. Additionally, the anatomic position markers may become mispositioned and could potentially be associated with the opposite side of the anatomy. (Added 1/3/2023), Expansion of the root cause to include incorrect amplimat field selection by 90 degrees following the first exposure using Wallstand VS2 and a subsequent system restart. Previously, this issue only occurred following a system restart where the Table Height 2 (TH2) is the default setting.
Root cause tag
software
FDA root cause
Software Design Change
Action
Philips notified customers of the issue via an Electronic Product Radiation Defect notification letter dated July 12, 2022. Philips will schedule an appointment with customers to install the software update at no cost to customers. For questions contact Philips Customer Care Solutions Center 1-800-722-9377.
Quantity in commerce
11 systems (updated 1/3/2023)
Distribution
Nationwide Distribution including Puerto Rico
Product codes and lots
Model Number 712204   Serial Number 210182 210187 220002 210049 210051 210097 210052 210053 210055 220020

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1563-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1563-2022Class IIOngoingFDA Mandated