TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model...
Cardiac Assist, Inc began a recall of TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or cardiopulmonary... on . FDA classified it as Class I and gives the reason as "Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data". As of , the recall status on record is Terminated, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-1763-2022
- Classification
- Class I
- Status
- Terminated
- Initiated
- Posted
- Terminated
- Recalling firm
- Cardiac Assist, Inc
- Product
- TANDEM LIFE LivaNova LIFESPARC Controller; Controller and Accessories Model 5900-0000; Controller Only Model 5900-0001. For circulatory or cardiopulmonary bypass.
- Product code
- DWA Control, Pump Speed, Cardiopulmonary Bypass
- Reason for recall
- Critical Failure of the LifeSPARC controller when the software freezes or crashes and the screen does not display data.
- Root cause tag
- software
- FDA root cause
- Under Investigation by firm
- Action
- On July 21, 2022, the recalling firm notified impacted customers through mailed letters. Customers were instructed to use Operations Manual revision 15 and to discard all previous revisions of the Operations Manuals.
- Quantity in commerce
- 723 units (484 currently in field)
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Model 5900-0000 - UDI-DI: 00814112020562 Model 5900-0001 - UDI-DI: 00814112020623 Serial No.: C00115, C50232, C50253, C50403, C00195, C50619, C50623, C50161, C50167, C50388, C50432, C50505, C00189, C50437, C50671, C50690, C50362, C50648, C00048, C00087, C00149, C50274, C50506, C50308, C50319, C00040, C00107, C50160, C50016, C50604, C50493, C50355, C00126, C00155, C50113, C00041, C00135, C00042, C00035, C50012, C50019, C00176, C50386, C00052, C00076, C00092, C50490, C00095, C50149, C00182, C50392, C50337, C50451, C50306, C50460, C50512, C50082, C50092, C50174, C00151, C50328, C50349, C50111, C50501, C00157, C50363, C50353, C50476, C50212, C50219, C50350, C00174, C00006, C50076, C50105, C50235, C50613, C50197, C50375, C50463, C50517, C50632, C50662, C00199, C00205, C50699, C50269, C50398, C50618, C50645, C50100, C50101, C50626, C50204, C50266, C50330, C50031, C50050, C50241, C50524, C50181, C50365, C00038, C50237, C00190, C50598, C50005, C50213, C50214, C50283, C00143, C50287, C50288, C50259, C50285, C50270, C50639, C00158, C50357, C50601, C50289, C50059, C50065, C50043, C50037, C50072, C50096, C50153, C50140, C50703, C50391, C50536, C50682, C50314, C50354, C50470, C50409, C50526, C50146, C50225, C50226, C50001, C50090, C00169, C00023, C00161, C50224, C50230, C50254, C50458, C50525, C50581, C50591, C50378, C50428, C50466, C50539, C50303, C00063, C00071, C50211, C50229, C50280, C00167, C50374, C50387, C00141, C00136, C00145, C50312, C00073, C50243, C00150, C50222, C50343, C50425, C00053, C00057, C50652, C00113, C00114, C50038, C50247, C50651, C50521, C00072, C50485, C50492, C50057, C50126, C00152, C50530, C00086, C00094, C00181, C50389, C50399, C50453, C50169, C50148, C00148, C50456, C50015, C00098, C50465, C50448, C00179, C50406, C50026, C50044, C50162, C00016, C50056, C50170, C50042, C50058, C00039, C00078, C00093, C00089, C50157, C50234, C50236, C50318, C50395, C00166, C50327, C50356, C50410, C50471, C50262, C50385, C50195, C50196, C00132, C00130, C50381, C50436, C50435, C50047, C50055, C50064, C50429, C50370, C50377, C50486, C50627, C50664, C50066, C50078, C50468, C50093, C00129, C50186, C50338, C50483, C50216, C50077, C50108, C00180, C50414, C00079, C00172, C50248, C50257, C50592, C50640, C00083, C50041, C50364, C50452, C50074, C50479, C50575, C00201, C50046, C50326, C50498, C50430, C50438, C50607, C50018, C50175, C50014, C00144, C50477, C50418, C50267, C50119, C00128, C50024, C50185, C50256, C50290, C00165, C50400, C50508, C00015, C00044, C00045, C50344, C50631, C50649, C50635, C50182, C50198, C50251, C50503, C50084, C50122, C50130, C50115, C50228, C00184, C50507, C50087, C50532, C50188, C50104, C50531, C00051, C50051, C50123, C50294, C50444, C50600, C50271, C50272, C50296, C50313, C50279, C00138, C50194, C00054, C00081, C50376, C50624, C50630, C00171, C50434, C50457, C00175, C50599, C50263, C50497, C50496, C50494, C50322, C50411, C50107, C00146, C50089, C50473, C50176, C50307, C00194, C50171, C50459, C50616, C50636, C50647, C00085, C00137, C00139, C50298, C50348, C50215, C50351, C50667, C50663, C50641, C50120, C50109, C50537, C50145, C50514, C00109, C50035, C50010, C50080, C00046, C50393, C50397, C50446, C50462, C50644, C50657, C50656, C50439, C50449, C50334, C50052, C50062, C50190, C50300, C00163, C00164, C00043, C00082, C50033, C00156, C00080, C50070, C50003, C50004, C50020, C50191, C50187, C50405, C50422, C50511, C50116, C50184, C50143, C50154, C50060, C50063, C50299, C50379, C50488, C50331, C00183, C50136, C50139, C50177, C50141, C00178, C50582, C00088, C50025, C50193, C50495, C50007, C50110, C50164, C00168, C00050, C50221, C50489, C50596, C00206, C50685, C50584, C50673, C50324, C50345, C50382, C50320, C50472, C50474, C50183, C50205, C50323, C50440, C00192, C50629, C50579, C50646, C50661, C50589, C50321, C50424, C50192, C50220, C50252, C50402, C00066, C00101, C50481, C50029, C50217, C50233, C50467, C00049, C00062, C00200, C00204, C00068, C00034, C50255, C50342, C50309, C50361, C50006, C50048, C50036, C50068, C50401, C50333, C50081, C50049, C50103, C50346, C50347, C50519, C50152, C50297, C50369, C50373, C50189, C50292, C50316, C50528, C50620, C50246, C50242, C00056, C00077, C00160, C50595, C50594, C50311, C50304, C50121, C50454, C50239, C50273, C50282, C50416, C50621, C50504, C50441, C50587, C00154, C50227, C50368, C50478, C00170, C00173, C50339, C00153, C00159, C50264, C50396, C00055, C00105, C50284, C50605, C50132, C50583, C50231, C50240, C50461, C50352, C50383, C50527, C00112, C50028, C50013, C50159, C50455, C50295, C00074, C50201
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1763-2022 | Class I | Terminated | Voluntary: Firm initiated |