Z-1751-2022Class II

iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the...

Alcor Scientific LLC / initiated 2022-07-20 / Open, Classified / root cause: manufacturing process

Alcor Scientific, Inc. began a recall of iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to... on . FDA classified it as Class II and gives the reason as "Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-1751-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Alcor Scientific LLC
Product
iSED Primary Pump Tubing Spare Part Kit- (IVD) tubing is used to move sample and wash solution from the aspiration needle, through the reading chamber, and to the waste on the iSED Automated ESR Analyzer PN: 112-13-002
Product code
GKB Device, Automated Sedimentation Rate
Reason for recall
Primary Pump Tubing Spare Part is incorrect. If the tubing is installed onto an ALCOR iSED ESR Analyzer, there is a potential of a biohazard leak from the iSEDs Probe Tube
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Alcor Scientific notified consignees via Emails on 21 July 2022 that included an Advisory Notice (AN-001). Letter states reason for recall, health risk and action to take: " Remove from your inventory any Lot 4/22 Primary Pump Tubing Spare Part Kits (112-13-002). Destroy this tubing. " If the tubing has been installed on any iSED ESR Analyzers, the tubing must be immediately replaced with new Primary Pump Tubing (112-13-002) from any other lot and Alcor must be notified of the locations where that tubing was installed. " Complete this form and return via email to [email protected] within 3 days. ALCOR will replace all Primary Pump Tubing from Lot 4/22 that was shipped to your facility.
Quantity in commerce
38 units
Distribution
US Nationwide Distribution - FL, MA, MD, PA, SD, WI Foreign: Croatia Honduras Korea
Product codes and lots
Lot 4/22

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1751-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1751-2022Class IIOngoingVoluntary: Firm initiated