Z-1813-2022Class II

Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An...

OLYMPUS Corporation of the Americas / initiated 2022-08-12 / Open, Classified

Olympus Corporation of the Americas began a recall of Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally... on . FDA classified it as Class II and gives the reason as "An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1813-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
OLYMPUS Corporation of the Americas
Product
Olympus POWERSEAL 5MM, 23CM, CURVED JAW SEALER & DIVIDER, DOUBLE-ACTION Model Number: PS-0523CJDA. An electrosurgical device use in laparoscopic/minimally invasive or open surgical procedures.
Product code
GEI Electrosurgical, Cutting & Coagulation & Accessories
Reason for recall
An out of specification component preventing the blue retention ring from keeping the rotation knob in place. This may result in a loose rotation collar and retention ring and cause delay of treatment
FDA root cause
Under Investigation by firm
Action
Olympus issued Urgent Medical Device Recall Letter on 8/12/22 to Surgical Department Risk Management. Letter states reason for recall, health risk and action to take: 1. Immediately locate any product with affected lot numbers listed in this communication. See Attachment 2 for pictures showing the location of lot numbers on the package. 2. Call your Olympus customer service representative at 1 800 848 9024, option 2. Olympus will issue a Return Material Authorization to return any affected product at no charge to you. Olympus will issue a credit to your facility upon return of affected product. 3. Access the Olympus recall portal to indicate that you have received this notification. Go to https://olympusamerica.com/recall. Enter the recall number 0412 and provide your contact information as indicated in the portal. 4. If you have distributed these devices outside your facility, please notify your customers of this removal action immediately. Your notification to your customers may be enhanced by including a copy of this recall notification letter. Contact directly at (647) 999-3203 or at [email protected] for any additional information or support concerning this matter.
Quantity in commerce
27 units (5/box) US; 83 units(5/box) OUS
Distribution
Nationwide OUS: AUSTRALIA, CANADA , HONG KONG, CHINA, KOREA
Product codes and lots
UDI-DI: 00821925044531 Lot Numbers: CA191167 CA191168 CA191169 CA191170

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1813-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1813-2022Class IIOngoingVoluntary: Firm initiated
Z-1814-2022Class IIOngoingVoluntary: Firm initiated
Z-1815-2022Class IIOngoingVoluntary: Firm initiated