t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ...
Tandem Diabetes Care Inc began a recall of t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ... on . FDA classified it as Class II and gives the reason as "Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-1812-2022
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Tandem Diabetes Care, Inc.
- Product
- t:slim X2 insulin pump; t:slim X2 insulin pump with Dexcom G5; t:slim X2 insulin pump with Basal-IQ technology; t: slim x2 insulin pump with the Control-IQ Technology
- Product code
- PQF Sensor, Glucose, Invasive, Non-Adjunctive
- Reason for recall
- Insulin pumps may have the following issues: 1) Malfunction 6 Non-Volatile Memory, 2) Inaccurate (Fluctuating) Battery Life Display, 3) Touchscreen Staying On, 4) Unexpected Open Loop, that can be mitigated with a software update. Issues could result in hypoglycemia, hyperglycemia or diabetic ketoacidosis.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On 5/24/22, and 7/7/22, correction notices sent to customers in Australia and New Zealand to update to the latest software. t:slim X2 insulin pump with Basal-IQ technology version 6.6 and Control-IQ technology version 7.6 available with latest software versions. Software update contact: Australia, AMSL: [email protected] or 1300 851 056; New Zealand, NZMS [email protected] (Questions: 0508 634 103) On 7/26/22, correction notices were mailed or emailed in the United States to healthcare providers and distributors. Recalling firm contact: [email protected] Providers asked to instruct patients to update pump software. Software update called version 6.6.0.1 for Basal-IQ technology and version 7.6.0.1 for Control-IQ technology. Check software version on the pump: options > my pump > pump info and arrow down to software version. See user guide and to visit https://support.tandemdiabetes.com/hc/en-us/articles/1500011452302-How-do-I-check-what-software-version-is-on-my-pump-video- Latest software can be downloaded remotely from a computer. Visit https://portal.tandemdiabetes.com and login with their t:connect account credentials. Once logged in, click the MY DEVICES to view available software updates. Perform software update via the Tandem Device Updater (TDU), with precautions and instructions provided in user TDU guide: https://www.tandemdiabetes.com/docs/default-source/product-documents/tandem-device-updater/aw-1004796_c-user-guide-tandem-device-updater-5-4-artwork.pdf. Acknowledge correction notice receipt: https://go.tandemdiabetes.com/Field-Correction-Notice-X.6_US.html Distributors with impacted part numbers asked to contact the recalling firm s distribution team. On 8/4/22, users sent correction notice to update pump software. See above for acknowledgment form link and software update instructions. Prescription and training may be required if pump is out-of-warranty and not currently using Control-IQ technology, can only update to Basal-
- Quantity in commerce
- 8399
- Distribution
- US: NC, OK, ME, NY, TX, LA, NE, FL, AZ, IL, OH, CA, AL, MI, UT, VA, MA, SC, MS, GA, NV, HI, MN, WA, IN, AR, ID, CO, IA, MO, MT, PA, WI, NH, MD, WV, KS, NM, OR, KY, DE, ND, CT, NJ, TN, WY, AK, SD, DC, RI, IA, PR, VT, PT, VI, NC. OUS: Australia, New Zealand, South Africa
- Product codes and lots
- Insulin Pump/Software Versions: t:slim X2 insulin pump/6.4.1 and earlier; t:slim X2 insulin pump with Dexcom G5/5.2.1, 5.2.2; t:slim X2 insulin pump with Basal-IQ technology/6.3.0.1, 6.4, 6.4.1; t: slim x2 insulin pump with the Control-IQ Technology/7.4, 7.4.3
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-1811-2022 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-1812-2022 | Class II | Ongoing | Voluntary: Firm initiated |