BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting...

Becton Dickinson & Co / initiated 2022-09-15 / Open, Classified / root cause: design

Becton Dickinson & Company began a recall of BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting, separating, and processing blood for... on . FDA classified it as Class II and gives the reason as "The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.

Recall record

Recall number
Z-0095-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Becton Dickinson & Co
Product
BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg, Catalog Number 367001. For collecting, transporting, separating, and processing blood for blood alcohol determination.
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
The affected product contains isobutylene which has recently demonstrated potential interference in determination of methanol by gas chromatography methods. The interference may lead to false positive results in methanol testing.
Root cause tag
design
FDA root cause
Component design/selection
Action
On September 15, 2022, the firm notified affected distributors and customers of the recall via Urgent Medical Device Product Advisory letters. Customers were instructed to do the following: 1. Inspect your inventory for the affected catalog number. 2. Ensure the contents of the Product Advisory are read, understood and shared with laboratory management staff having oversight over gas chromatography testing for methanol. Review internal method validations, and if applicable, adjust test practices or settings to avoid potential interference (e.g., implementing the tube as a blank/control, adjustment of GC column and/or operational settings to prevent co-elution, or conversion to dual-column system). 3. If the potential interference cannot be mitigated, stop using the BD Vacutainer Tube Sodium Fluoride: 100 mg, Potassium Oxalate: 20 mg. Product may be returned for credit. The recalling firm will update the IFU to address the potential risk of interference from rubber stopper outgassing in gas chromatography methods. The firm has also initiated corrective and preventive action to investigate root cause and identify actions. If you have further questions, please contact the firm via Phone: 1-844-8BD-LIFE (1-844-823-5433) Say Recall when prompted Mon Fri 8:00am and 5:00pm CT or Email: [email protected]
Quantity in commerce
2,196,000
Distribution
Distribution US nationwide, Canada, New Zealand, and Hong Kong.
Product codes and lots
UDI-DI: Case: 50382903670012; Shelf: 30382903670018; Each: 00382903670017; ALL LOTS are affected.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0095-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0095-2023Class IIOngoingVoluntary: Firm initiated