Z-0133-2023Class II

da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC)...

Intuitive Surgical, Inc. / initiated 2022-09-20 / Open, Classified / root cause: software

Intuitive Surgical, Inc. began a recall of da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS) on . FDA classified it as Class II and gives the reason as "Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3)...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.

Recall record

Recall number
Z-0133-2023
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Intuitive Surgical, Inc.
Product
da Vinci X (IS4200) and Xi (IS4000) systems, which consists of finished devices Surgeon Console (SSC), Patient Cart (PSCART), Vision System (VSS)
Product code
NAY System, Surgical, Computer Controlled Instrument
Reason for recall
Inadvertent energy delivery from surgical system instrument if 1) Force bipolar and bipolar instruments installed on system 2) Force bipolar connected to Force Triad generator, bipolar not connected to generator 3) Yellow pedal associated with force bipolar is pressed, released with head in surgeon console 4) Blue pedal associated with bipolar is pressed, resulting in force bipolar energy delivery
Root cause tag
software
FDA root cause
Software Design Change
Action
On 09/20/22, correction notices were mailed to customers. Customers were advised of the following: 1) The system software version can be located on the Vision Cart Touchscreen in the Troubleshooting Tab. 2) Ensure all device users read and understand this notice. 3) Notice should be passed on to all those who need to be aware within your organization or to any organization where the potentially affected devices have been transferred. 4) Ensure proper instrument connections to the generator(s) and continue to follow all existing warnings and cautions found in user manuals 551400-15, USER MANUAL,SYS,P10,US,IS4000, 553873-07, USER MANUAL,I-A,MDR P10,IS4000-IS4200 and 554070-04, ADD,I-A,FORCE BIPOLAR,IS4000,IS4200. 5) Complete and return the acknowledgement form. 6) The recalling firm will follow-up with customers and schedule a software upgrade. 7) After the upgrade, inform all personnel that the software update/upgrade was completed. 8) If you have experienced this issue or require additional assistance, contact Technical Support: 800-876-1310, Option 1 (24/7) or [email protected]
Quantity in commerce
162
Distribution
US: MS, PA, WI, KY, TN, MI, FL, GA, NC, NV, LA, OH, IL, MO, IN, CA, AK, MD, TX, IA, MA, NE, AZ, WA, CO, AL, SC
Product codes and lots
Software version P10 (OS4 v10.0.0/A70_P10_B738) SOFTWARE,EMBEDDED RLS,IS4000 AND IS4200,A70_P10_B738 (Model: 610092-738) on Model Name/Model Number/UDI-DI: ASSY,PSCART,IS4000,4-ARM,P10/380652-55/00886874110720, ASSY,SSC,IS4000,P10/380677-28/00886874110744, ASSY,VSS VISION SYSTEM,IS4000,P10/381121-45/00886874110898, ASSY,PSCART,IS4200,P10/380620-47/00886874115404, ASSY,VSC,IS4000,RECONDITIONED,P10/380721-15/00886874115343, ASSY,SSC,IS4000,RECONDITIONED,P10/380723-16/00886874115374

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0133-2023
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0133-2023Class IIOngoingVoluntary: Firm initiated