Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-1761-2022Class II

NanoSight LM10 fitted with LM12 module. A laboratory instrument for determination of particle characteristics...

Malvern Panalytical Ltd / initiated 2022-02-28 / Open, Classified

MALVERN PANALYTICAL LTD began a recall of NanoSight LM10 fitted with LM12 module. A laboratory instrument for determination of particle characteristics (e.g. size) on . FDA classified it as Class II and gives the reason as "Product found to not comply to Class 1 Laser safety requirement". As of , the recall status on record is Open, Classified.

Recall record

Recall number
Z-1761-2022
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Malvern Panalytical Ltd
Product
NanoSight LM10 fitted with LM12 module. A laboratory instrument for determination of particle characteristics (e.g. size).
Product code
REN Particle-Size Measuring Instrument
Reason for recall
Product found to not comply to Class 1 Laser safety requirement.
FDA root cause
Radiation Control for Health and Safety Act
Action
Customers are being made aware of this issue. Malvern Panalytical will retrofit modified microscope plates and cell tops with a new safety cover on the LM10 to maintain the original Class 1 laser product classification at no charge to the purchaser.
Quantity in commerce
111 units
Distribution
US Nationwide Distribution
Product codes and lots
CDRH accession # 0510530-001

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-1761-2022
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-1761-2022Class IIOngoingFDA Mandated
Z-1762-2022Class IIOngoingFDA Mandated